Programmable LCD Artificial Climate Chamber, 450 L
Research-grade laboratory product
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||
|
LABORATORY PROCUREMENT SUPPORT
For model selection, accessory matching, installation-space review, power-supply confirmation or configuration support, contact our technical sales team.
E-MAIL: inquiry@atomfair.com
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Supplier: Atomfair
Brand: ATOMFAIR®
|
The chamber requires a controlled environment of +5 to 30°C and a dedicated AC 220V 50Hz power supply. Continuous operation exceeding 180 hours requires proper installation and monitoring of the independent over-temperature alarm.
- Ambient Temperature Requirement: The working environment must be maintained between +5 and 30°C to ensure proper temperature control and system stability.
- Power Supply Specification: The chamber requires a dedicated AC 220V 50Hz power supply with a capacity of 2050 W to avoid electrical overload or instability.
- Continuous Operation Guidance: Continuous operation is recommended for at least 180 hours only under proper installation conditions with adequate ventilation and alarm monitoring.
- Installation Clearance: Maintain accessibility to air inlets, test ports, shelving, drain areas, doors, and controller surfaces for cleaning and maintenance.
- Over-Temperature Alarm: The independent over-temperature alarm must be verified functional before unattended running to protect samples from thermal excursions.
Unpack and position the chamber following the manufacturer's guidelines for lifting and clearance. Connect the power supply and configure the temperature, humidity, and light parameters using the LCD controller before loading samples.
Required Equipment: Suitable lifting equipment (e.g., forklift or pallet jack), LCD microcomputer controller interface
- Unpack and Position
Unpack the chamber and keep it upright, using suitable lifting equipment to position it near the installation site. - Confirm Installation Space
Confirm that the installation space accommodates the external dimensions of 950 x 850 x 1850 mm with adequate clearance. - Connect Power Supply
Connect the chamber to a dedicated AC 220V 50Hz power supply that can supply the required 2050 W load. - Maintain Access Points
Keep all air inlets, test ports, shelving, drain areas, doors, and controller surfaces accessible for cleaning and maintenance. - Configure and Load Samples
Set the desired temperature, humidity, and light intensity values using the LCD controller menu and load samples onto the removable shelves.
Why does the MGC-450HP artificial climate chamber have a narrower temperature range when the lights are on (10 to 50°C) compared to no-light mode (4 to 50°C)?
The illuminated mode range is limited to 10–50°C because the three-side lighting system (0–25000 lx, six-step adjustable) generates internal heat that raises the minimum achievable temperature. In no-light mode the compressor can reach 4°C without this thermal load. The specification tables for the AF-MHP0450 model list both ranges and confirm a 6°C higher minimum when illumination is active.
Can the AF-MHP0450 climate chamber be configured for CO₂ enrichment in plant physiology studies?
Yes, CO₂ support is available as an optional accessory. The chamber can be fitted with a CO₂ inlet and an infrared CO₂ controller for monitoring and regulation. These options are listed under ‘Optional Accessories and Functions’ and must be specified at the time of ordering; they are not included in the base configuration.
Why does the MGC-450HP chamber recommend a minimum continuous operation of 180 hours?
The manufacturer specifies a recommended continuous operation schedule of not less than 180 hours under proper installation conditions to maintain stable temperature and humidity control. This guideline helps prevent performance drift in the imported compressor and humidity-sensor system during long-term cultivation or simulation runs. Extended off-cycles could compromise reproducibility in humidity-sensitive studies.
The Atomfair MGC-450HP (SKU AF-LE-E-MGAC-0450-HP45) is a 450-L artificial climate chamber combining temperature, humidity, and programmable light control with a stainless-steel interior, suitable for plant growth and environmental simulation. Its ±1°C temperature uniformity and ±5–7%RH humidity tolerance define operational boundaries for precision work.
Positive
- Integrated humidity and illumination control: Coordinated temperature, humidity, and light cycling with six-step adjustable illumination up to 25,000 lx supports complex plant cultivation and environmental simulation protocols.
- Stainless steel chamber with removable shelving: 304 stainless-steel interior facilitates cleaning and decontamination between experiments, reducing cross-contamination risk in long-duration studies.
Trade-offs
- Power and space requirements: Requires a 2050 W, 220 V, 50 Hz supply and external footprint of 950 x 850 x 1850 mm, necessitating dedicated lab infrastructure and adequate floor space.
- Humidity control tolerance: Humidity deviation is ±5–7% RH, which may limit applicability for assays demanding tighter humidity stability.
Every advanced material, component, equipment, and instrument in our catalog is backed by rigorous testing. We maintain strict internal quality management frameworks and align with CE conformity metrics to deliver transparent, reproducible performance data via our public open-science repository.
To request raw batch performance data, submit formal vendor registration paperwork, or execute a fast-turnaround R&D manufacturing loop, contact us at inquiry@atomfair.com.
Item is dispatched under the Atomfair Shipping & Delivery Framework (Free worldwide shipping on orders over $59 USD excl. heavy equipment). Return is governed by the Atomfair Return & Refund Policy (7-day technical return window).