Color Touchscreen Constant Temperature and Humidity Chamber, 786 L,-10-100℃
Research-grade laboratory product
|
|||||||||||||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||||||||||||
|
LABORATORY PROCUREMENT SUPPORT
For model selection, accessory matching, installation-space review, power-supply confirmation or configuration support, contact our technical sales team.
E-MAIL: inquiry@atomfair.com
|
|||||||||||||||||||||||||||||||||||||||||||||||||
|
Supplier: Atomfair
Brand: ATOMFAIR®
|
This chamber requires a stable indoor installation with a dedicated AC 380 V, 50 Hz power supply and adequate service clearance. The unit must remain upright during transport, and all alarm conditions must be promptly addressed to prevent equipment damage or sample loss.
- Installation Space: The chamber requires a minimum installation space of 860 x 1340 x 1995 mm with service access around all sides.
- Power Supply: The chamber must be connected to an AC 380 V, 50 Hz power supply rated for 4550 W with appropriate overcurrent and grounding.
- Transport Orientation: The chamber must remain upright during transport, unpacking, and positioning to prevent compressor damage.
- Accessibility: All air inlets, test ports, shelving, drain areas, doors, and controller surfaces must remain accessible for cleaning and maintenance.
- Alarm Response: Operator must respond to over-temperature, overload, over-current, compressor overheating, fan overheating, and water-shortage alarms without delay.
This procedure covers the unpacking and installation of the 786 L constant temperature and humidity chamber. Follow these steps to ensure safe positioning and power supply preparation.
Required Equipment: Suitable lifting equipment (e.g., forklift or pallet jack)
- Confirm installation space and power supply
Confirm the installation space dimensions of 860 x 1340 x 1995 mm and prepare the specified AC 380 V, 50 Hz power supply. - Unpack and position the chamber
Unpack the chamber using suitable lifting equipment and position it upright on a level surface. - Verify accessibility for maintenance
Verify that all air inlets, test ports, shelving, drain areas, doors, and controller surfaces are accessible for cleaning and maintenance.
What temperature uniformity can I expect from the BH-800CL chamber at high vs low temperature setpoints?
The chamber specifies temperature fluctuation/uniformity of ±0.5°C at high temperature (above 0°C) and ±1°C at low temperature (below 0°C), based on the technical specifications table. This performance is supported by the PID control and imported sealed-compressor refrigeration system.
What remote monitoring and data logging options are available for the Atomfair BH-800CL chamber?
The chamber supports optional RS485 communication interface and software, a wireless alarm system, and an App control system for remote status review and alarm notification. Additionally, an embedded or thermal printer can be added for local data logging. These options help preserve records and monitor chamber status away from the instrument.
What are the installation requirements for the BH-800CL chamber, including power and space?
The chamber requires a stable indoor laboratory installation with service access around the unit. It needs an AC 380 V, 50 Hz power supply with a rated input of 4550 W. The external dimensions are 860 x 1340 x 1995 mm, and the working environment must allow access to air inlets, test ports, shelving, drain areas, doors, and controller surfaces for cleaning and maintenance.
The AF-LE-E-BHHC-0800-CL80 (786 L, -10 to 100 °C, 10–95% RH) uses an imported sealed-compressor refrigeration system with PT100 and capacitive humidity sensors, providing ±0.5 °C high-temperature fluctuation and ±3% RH deviation for controlled humidity stability studies. The 4550 W power requirement at 380 V and 860 x 1340 x 1995 mm footprint demand dedicated lab infrastructure.
Positive
- Capacitive humidity sensor avoids wet-bulb maintenance: Imported capacitive electronic humidity sensor provides continuous real-time monitoring without wet-bulb wick replacement, reducing routine maintenance in humidity-sensitive incubation and stability-study workflows.
- Color touchscreen with 100-program capability: The programmable controller supports 100 programs, 1000 segments, and 999 cycle steps with multilingual interface and live curve display, enabling complex staged methods for conditioning and testing protocols.
Trade-offs
- Requires 380 V three-phase power supply: The 4550 W input demands AC 380 V, 50 Hz power, which may not be standard in all laboratory settings and requires dedicated electrical infrastructure planning.
- Large footprint with service access constraints: External dimensions of 860 x 1340 x 1995 mm require stable indoor installation with clearance for air inlets, test ports, and drain areas, limiting placement flexibility in crowded labs.