Programmable Touchscreen Constant Temperature and Humidity Chamber, 500 L,-10-100℃
Research-grade laboratory product
|
|||||||||||||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||||||||||||
|
LABORATORY PROCUREMENT SUPPORT
For model selection, accessory matching, installation-space review, power-supply confirmation or configuration support, contact our technical sales team.
E-MAIL: inquiry@atomfair.com
|
|||||||||||||||||||||||||||||||||||||||||||||||||
|
Supplier: Atomfair
Brand: ATOMFAIR®
|
The chamber requires a stable +5 to 30°C working environment and a dedicated AC 220V, 50Hz power supply rated at 3850W. Operate within -10 to 100°C and 35% to 95%RH ranges; the unit includes over-temperature, overload, over-current, fan, and compressor protection systems.
- Power Supply Constraint: The unit must be connected to a dedicated AC 220V, 50Hz power supply with sufficient capacity for the 3850W load to avoid over-current conditions.
- Environmental Range: The chamber is designed for operation within +5 to 30°C ambient temperature and 35% to 95%RH humidity, with temperature control from -10 to 100°C.
- Cooling and Refrigeration: The balancing-control refrigeration system includes multi-point safety protection and requires unrestricted airflow around the unit for proper heat rejection.
- Chamber Material: The imported 304 mirror-finish stainless-steel interior supports repeated cleaning and hygiene protocols but may be damaged by abrasive or corrosive cleaning agents.
- Safety Protections: Independent alarm and protection functions cover over-temperature, overload, over-current, fan, and compressor conditions to prevent equipment damage and sample loss.
Unpack and position the chamber upright in the prepared installation space of 850 x 1100 x 1930 mm. Connect to the specified AC 220V, 50Hz power supply and verify all safety protections before operation.
Required Equipment: Lifting equipment
- Confirm Installation Space
Confirm that the installation space measures at least 850 x 1100 x 1930 mm and ensure adequate clearance for air inlets, test ports, and drain areas. - Prepare Power Supply
Prepare a dedicated AC 220V, 50Hz power supply with sufficient capacity to handle the 3850W load and install appropriate over-current protection. - Unpack and Position Chamber
Keep the shipped chamber upright during unpacking and use suitable lifting equipment to move and position the unit at the prepared location. - Inspect Chamber Access Points
Inspect all test ports, shelving, door seals, and controller surfaces for accessibility and cleanliness before connecting power. - Verify Safety Protections
Verify that the multi-point safety protection system, including over-temperature, overload, over-current, fan, and compressor alarms, is functional before starting the unit.
What is the temperature uniformity difference between the high and low temperature ranges in the Atomfair BPS-500CL chamber?
The temperature fluctuation/uniformity is ±0.5°C at high temperature and ±1°C at low temperature, as specified in the technical data. This difference means the chamber achieves tighter control at elevated temperatures than at sub-zero conditions, which is critical for applications requiring precise thermal stability across the -10 to 100°C range.
What are the critical installation infrastructure requirements for the 500 L constant temperature and humidity chamber?
The chamber requires a 3850 W power supply at AC 220 V, 50 Hz, and an ambient working environment of +5 to 30°C. The external dimensions are 850 x 1100 x 1930 mm, so adequate clearance must be provided for ventilation and maintenance access. Confirm the installation space and power supply before ordering.
How does the balancing-control refrigeration system with multi-point safety protection affect long-term reliability?
The balancing-control refrigeration system is designed for stable low-temperature performance, and multi-point safety protection covers over-temperature, overload, over-current, fan, and compressor conditions. This configuration helps maintain dependable chamber recovery during repeated use and reduces the risk of sample loss due to system failure.
The AF-LE-E-BPSC-0500-CL50 (BPS-500CL) is a 500 L programmable constant temperature and humidity chamber with -10 to 100°C range and 35–95%RH control, featuring a PID auto-tuning touchscreen controller, 304 stainless steel interior, and balancing refrigeration. It is suited for long-duration incubation, stability testing, and climate simulation but requires a 3850 W 220V/50Hz supply and a conditioned laboratory environment (+5–30°C) for reliable operation.
Positive
- Large program capacity for complex protocols: Supports up to 100 programs, 1000 segments, and 999 cycles with per-segment timing up to 99 h 59 min, enabling multi-stage environmental test methods without external sequencing.
- PID auto-tuning and balancing refrigeration: PID self-tuning and balancing-control refrigeration system maintain temperature and humidity stability, reducing deviation in humidity-sensitive incubation and conditioning applications.
Trade-offs
- High power and electrical requirements: Rated 3850 W at AC 220V/50 Hz, requiring a dedicated circuit and limiting compatibility with labs using 60 Hz or 110V supplies without a transformer.
- Restricted ambient operating range: Specified working environment of +5 to 30°C restricts placement to climate-controlled laboratory spaces; operation outside this range may degrade performance or trigger protection.
Every advanced material, component, equipment, and instrument in our catalog is backed by rigorous testing. We maintain strict internal quality management frameworks and align with CE conformity metrics to deliver transparent, reproducible performance data via our public open-science repository.