Description
ATOMFAIR® LOW TEMPERATURE FREEZING TEST CHAMBERRESEARCH GRADE EQUIPMENT
|
|||||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||||
|
TAILORED SOLUTIONS FOR RESEARCH
Contact our engineering team for technical support or official quotations.
EMAIL: inquiry@atomfair.com
|
|||||||||||||||||||||||||||||||||||||||||
|
Manufacturer: Atomfair LLC
Brand: ATOMFAIR®
|
Operate within specified environmental limits: ambient 0-23°C, relative humidity ≤75%, in a well-ventilated dry area free of corrosive gases, direct sunlight, water sources, heat sources, and magnetic fields. Maintain minimum clearances of 50cm from the rear wall and 15cm from side walls; use an independent grounded socket rated ≥10A with a voltage stabilizer if supply is unstable.
- Electrical Safety: Ensure the unit is connected to a dedicated grounded power socket with current capacity ≥10A and that voltage is stabilized to 220V/50Hz/15A if line voltage fluctuates.
- Environmental Restrictions: Avoid operation in damp environments or atmospheres containing corrosive gases, and keep the unit away from direct sunlight, water sources, heat sources, magnetic fields, and volatile or corrosive substances.
- Material Storage Prohibition: Do not store volatile, flammable, explosive materials, strong acids, strong alkalis, gasoline, or alcohol inside the chamber.
- Operational Constraints: Supervise the unit during operation, turn off power immediately during a power failure, and allow a minimum 30-minute interval before restarting to prevent compressor damage.
- Transport and Setup: Maintain a tilt angle ≤45° during transportation and remove all packaging before use; after a long idle period, run the unit for 2 hours before testing to ensure accurate data.
Follow this procedure to safely start, load, and shut down the chamber, and to perform routine maintenance. Ensure the unit is properly grounded and placed in a compliant environment before proceeding.
Required Equipment: Independent grounded power socket (≥10A), Voltage stabilizer (5x input power capacity, if voltage unstable), Soft cloth and mild soapy water for cleaning
- Pre-operation Check
Confirm the power socket is independent and grounded, voltage is 220-240V at 49-51Hz, and the panel power switch is in the OFF position. - Startup Procedure
Plug in the power cord, turn the panel power switch to ON, and allow the compressor to start after the built-in time delay. - Load Samples
After the chamber temperature reaches the set value, place samples evenly with gaps for air circulation, wrapping high-humidity or easily dehydrated samples in sealed bags. - Monitor During Operation
Supervise the unit continuously and do not overload with excessive samples to avoid continuous compressor operation and slow cooling. - Shutdown Procedure
Turn off the panel power switch first (display off), then disconnect the external power supply. - Cleaning and Defrosting
After testing, remove samples, open the door for natural defrosting, drain water, and wipe the inner chamber dry with a soft cloth. - Maintenance
Clean the door gasket gently with mild soapy water and avoid using boiling water, acidic agents, or chemical thinners on any part of the unit.
How does the ambient operating temperature range of 0-23°C affect the AFDW5030L's ability to achieve and maintain -50°C?
The AFDW5030L is designed to operate within an ambient temperature of 0-23°C. Operating outside this range can degrade cooling performance and may prevent the chamber from reaching its minimum -50°C or maintaining stable temperatures. The temperature resolution is 0.1°C, but ambient conditions directly influence compressor efficiency and thermal retention.
What sample types or chemical materials are incompatible with the AFDW5030L low-temperature chamber?
The AFDW5030L strictly prohibits storing volatile, flammable, explosive materials, strong acids, strong alkalis, gasoline, and alcohol due to safety hazards and potential damage. High-humidity or easy-dehydration samples must be sealed in moisture-proof wrapping to prevent frost accumulation that impairs cooling. The corrosion-resistant alloy inner tank is not rated for corrosive substances.
What are the required power supply specifications and restart intervals to avoid compressor failure in the AFDW5030L?
The AFDW5030L requires an independent grounded socket with current capacity ≥10A and stable voltage of AC 220-242V/50Hz. If voltage is unstable, a stabilizer of at least 5 times input power (adjusted to 220V/50Hz/15A) must be installed. After any power interruption, wait at least 30 minutes before restarting to prevent compressor damage; upon first use or after prolonged idle, run for 2 hours before loading samples.
The AFDW5030L is a 30L -50°C freezing chamber with a corrosion-resistant alloy inner tank and polyurethane insulation, suitable for low-temperature storage and testing of materials and biological samples. Its digital temperature control provides 0.1°C resolution, but it requires stable environmental conditions (0-23°C, ≤75% RH) and careful sample loading to maintain performance.
Positive
- Precise digital temperature control: The unit features a digital temperature control system with 0.1°C resolution, enabling accurate low-temperature setpoints down to -50°C for reproducible freezing tests and sample storage.
- Durable corrosion-resistant construction: The inner tank is made of corrosion-resistant alloy plate and the shell is high-quality steel with thickened polyurethane insulation, providing thermal retention and long-term resistance to aggressive sample environments.
Trade-offs
- Strict ambient and power requirements: Operational ambient temperature must be between 0-23°C with relative humidity ≤75%, and the unit requires a dedicated grounded socket with ≥10A capacity. Unstable voltage necessitates an external stabilizer of >5x input power to prevent compressor damage.
- Sample loading and maintenance constraints: Samples must leave gaps for air circulation, high-moisture items require sealed wrapping, and overloads must be avoided. After each run, the chamber requires defrosting and interior drying, and the unit must not be used with volatile, flammable, or corrosive materials.