The chamber must be operated within an ambient temperature range of +5 to 30°C to maintain specified cooling performance. Electrical supply must be a dedicated AC 220 V, 50 Hz circuit with capacity for the 4150 W peak load, and the unit must be positioned with unobstructed airflow to the refrigeration system.
- Power Supply: Connect the chamber to a dedicated AC 220 V, 50 Hz power supply with proper grounding and sufficient capacity for the 4150 W load.
- Ambient Temperature: Maintain the working environment between +5 and 30°C to avoid overheating or insufficient cooling and humidity control.
- Upright Transport: Keep the packed chamber upright at all times and use suitable lifting equipment during unloading and positioning.
- Accessibility: Ensure air inlets, test ports, shelving, drain areas, doors, and controller surfaces remain accessible for cleaning and maintenance.
Unpack and position the chamber in the designated installation space, then connect to the specified power supply. Program the touchscreen controller with the desired environmental parameters and allow the chamber to stabilize before loading samples.
Required Equipment: Lifting equipment (e.g., forklift or pallet jack)
- Inspect and Unpack
Inspect the chamber for any shipping damage before unpacking and keep the unit upright during removal of packaging. - Position the Chamber
Position the chamber in the designated installation space, ensuring adequate clearance for air inlets, test ports, and service access as per the external dimensions of 850 x 1100 x 1930 mm. - Connect Power Supply
Connect the chamber to a dedicated AC 220 V, 50 Hz power supply with proper grounding and verify the electrical connection meets the 4150 W requirement. - Program Environmental Parameters
Set the desired temperature, humidity, and program segments using the color touchscreen controller, utilizing up to 100 programs and 1000 segments. - Stabilize Chamber
Allow the chamber to stabilize at the set conditions, including PID auto-tuning if initiated, before loading any samples.
How does the temperature uniformity vary between high and low temperature setpoints in this chamber?
At high temperature setpoints, the temperature fluctuation and uniformity is ±0.5°C, while at low temperature setpoints it is ±1.0°C. This specification is directly stated in the product datasheet for the 500 L chamber.
What is the maximum program capacity of the touchscreen controller for staged environmental testing?
The programmable color touchscreen controller supports up to 100 programs, 1000 segments, and 999 cycle steps, with a maximum time setting of 99 hours and 59 minutes per segment. This allows complex, long-duration test protocols to be executed automatically.
What are the electrical supply and installation space requirements for the 500 L chamber?
The chamber requires an AC 220 V, 50 Hz power supply with a power input of 4150 W. External dimensions are 850 x 1100 x 1930 mm, and the working environment must be +5 to 30°C. Adequate clearance for air inlets, test ports, and maintenance access should be planned.
The Atomfair BPS-500CA (SKU: AF-LE-E-BPSC-0500-CA50) is a 500-liter programmable constant temperature and humidity chamber with a -20 to 100°C range and 35–95%RH, featuring a color touchscreen controller supporting 100 programs and PID auto-tuning, plus a 304 stainless-steel interior. Suitable for controlled environment studies, it requires a 4150 W, 220 V, 50 Hz power supply and an ambient temperature of 5–30°C.
Positive
- Programmable Touchscreen with Curve Display: The color touchscreen controller supports up to 100 programs, 1000 segments, and 999 cycles, with PID auto-tuning for stable temperature and humidity control, enabling staged environmental methods and repeatable laboratory workflows.
- 304 Stainless-Steel Chamber with Balancing Refrigeration: Imported 304 mirror-finish stainless-steel interior and balancing-control refrigeration system provide reliable long-duration operation, easy cleaning, and stable chamber recovery under repeated use.
Trade-offs
- Ambient Temperature Requirement: The chamber's specified working environment is +5 to 30°C, necessitating installation in a temperature-controlled laboratory and limiting use in unconditioned spaces.
- Electrical Power and Supply Dependency: The unit requires 4150 W at AC 220V 50 Hz, requiring dedicated power infrastructure and potentially limiting installation to facilities with compatible electrical systems.