Programmable Touchscreen Constant Temperature and Humidity Chamber, 50 L
Research-grade laboratory product
|
|||||||||||||||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||||||||||||||
|
LABORATORY PROCUREMENT SUPPORT
For model selection, accessory matching, installation-space review, power-supply confirmation or configuration support, contact our technical sales team.
E-MAIL: inquiry@atomfair.com
|
|||||||||||||||||||||||||||||||||||||||||||||||||||
|
Supplier: Atomfair
Brand: ATOMFAIR®
|
The chamber must be operated within an ambient temperature range of +5 to 30°C and connected to a dedicated AC 220 V, 50 Hz power supply. The unit requires upright positioning on a level surface with unobstructed air inlets, test ports, and drain areas to maintain safe operation.
- Ambient Temperature Range: Operate the chamber only within an ambient temperature range of +5 to 30°C.
- Electrical Supply: Connect the chamber to a dedicated AC 220 V, 50 Hz power supply rated at 1450 W.
- Upright Orientation: Keep the chamber upright during transport, unloading, and positioning to prevent damage.
- Accessibility Requirements: Ensure all air inlets, test ports, drain areas, doors, and controller surfaces remain accessible for cleaning and maintenance.
- Space Requirements: Confirm that the installation space accommodates the external dimensions of 720 x 620 x 725 mm.
This procedure outlines the steps to safely unpack, position, and connect the chamber for first-time use. Proper installation ensures reliable operation and extends equipment lifespan.
Required Equipment: Lifting equipment
- Position the chamber
Use suitable lifting equipment to position the chamber upright on a level surface that meets the external dimensions of 720 x 620 x 725 mm. - Prepare and connect power supply
Prepare and connect the dedicated AC 220 V, 50 Hz power supply rated at 1450 W. - Verify accessibility
After installation, verify that all air inlets, test ports, shelving, drain areas, doors, and controller surfaces are unobstructed.
Does the BPS-50CH constant temperature and humidity chamber provide active cooling or refrigeration?
No, this CH model does not include a refrigeration function. The temperature control range is RT+10 to 85°C, with a working environment of +5 to 30°C. This means the chamber can only cool to 15°C above ambient temperature at minimum. For applications requiring temperatures below ambient, a different model with refrigeration should be selected.
What data logging and remote monitoring options are available for the BPS-50CH chamber?
The chamber offers optional accessories including an embedded printer, RS485 communication interface with software, and a wireless alarm system. The programmable color touchscreen controller supports up to 100 programs, 1000 segments, and 999 cycle steps, with real-time curve display and PID auto-tuning. These options allow for data output, remote monitoring, and alarm notification, but must be specified at time of ordering.
What are the electrical installation requirements and safety protections for the BPS-50CH chamber?
The chamber requires AC 220V, 50Hz power supply with a rated input of 1450W. It includes independent alarm and protection functions for over-temperature, overload, over-current, fan, and compressor conditions. The external dimensions are 720 x 620 x 725 mm, and the working environment must be +5 to 30°C. Proper installation planning should ensure adequate space and power infrastructure.
The Atomfair BPS-50CH (SKU: AF-LE-E-BPSC-0050-CH50) is a 50L programmable touchscreen constant temperature and humidity chamber with a working range of RT+10 to 85°C and 80% to 95% RH, utilizing a balancing-control refrigeration system and imported 304 stainless-steel interior for durable, contamination-resistant operation. Its controller supports up to 100 programs with 1000 segments and PID auto-tuning, making it suitable for staged stability studies and incubation protocols in laboratory settings.
Positive
- Programmable Touchscreen and PID Control: The color touchscreen controller supports up to 100 programs, 1000 segments, and 999 cycles with PID auto-tuning, enabling precise, repeatable environmental profiles for multi-step protocols.
- Corrosion-Resistant 304 Stainless-Steel Interior: The imported 304 mirror-finish stainless-steel chamber interior minimizes contamination and facilitates repeated cleaning, essential for pharmaceutical and biotech applications requiring strict hygiene.
Trade-offs
- Limited Humidity and Temperature Range: Humidity control is restricted to 80%–95% RH and temperature spans only RT+10°C to 85°C, excluding low-humidity or sub-ambient conditioning without specified refrigeration support.
- Ambient Operating and Infrastructure Dependencies: The chamber requires a working environment of +5 to 30°C and a 220 V, 50 Hz power supply with 1450 W capacity, limiting placement to climate-controlled labs with dedicated electrical infrastructure.
Every advanced material, component, equipment, and instrument in our catalog is backed by rigorous testing. We maintain strict internal quality management frameworks and align with CE conformity metrics to deliver transparent, reproducible performance data via our public open-science repository.
To request raw batch performance data, submit formal vendor registration paperwork, or execute a fast-turnaround R&D manufacturing loop, contact us at inquiry@atomfair.com.
Item is dispatched under the Atomfair Shipping & Delivery Framework (Free worldwide shipping on orders over $59 USD excl. heavy equipment). Return is governed by the Atomfair Return & Refund Policy (7-day technical return window).