Balanced-Control Constant Temperature and Humidity Chamber, 50 L
Research-grade laboratory product
|
|||||||||||||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||||||||||||
|
LABORATORY PROCUREMENT SUPPORT
For model selection, accessory matching, installation-space review, power-supply confirmation or configuration support, contact our technical sales team.
E-MAIL: inquiry@atomfair.com
|
|||||||||||||||||||||||||||||||||||||||||||||||||
|
Supplier: Atomfair
Brand: ATOMFAIR®
|
The chamber requires a stable ambient temperature between +5 and 30°C and a dedicated AC 220 V, 50 Hz power supply. Independent over-temperature alarm must be enabled to protect samples during unattended operation.
- Ambient Temperature Range: Operate the chamber only in environments with ambient temperature between +5 and 30°C to maintain proper performance.
- Electrical Supply: Connect the chamber to a dedicated AC 220 V, 50 Hz power supply rated at 1450 W to avoid electrical hazards.
- Over-Temperature Alarm: Enable the independent over-temperature alarm before any unattended testing to automatically notify operators of abnormal conditions.
- Installation Orientation: Keep the chamber upright during transport and positioning, and use suitable lifting equipment to prevent tipping or damage.
- Air Inlet and Ventilation: Maintain clearance around air inlets, test ports, and drain areas to ensure proper airflow and prevent overheating.
Unpack and position the chamber upright, connect to the specified power supply, and configure the over-temperature alarm before use. Perform routine cleaning of the stainless-steel interior to maintain hygiene and performance.
Required Equipment: Lifting equipment
- Unpack and Position
Unpack the chamber and position it upright using suitable lifting equipment, ensuring the installation space accommodates the external dimensions of 720 x 620 x 725 mm. - Connect Power Supply
Connect the chamber to a dedicated AC 220 V, 50 Hz power source with a capacity of 1450 W. - Verify Ambient Conditions
Verify that the ambient temperature of the installation site is between +5 and 30°C to ensure proper operation. - Clear Air Inlets and Ports
Ensure air inlets, test ports, drain areas, and shelving are unobstructed and accessible for cleaning and maintenance. - Enable Over-Temperature Alarm
Enable the independent over-temperature alarm before starting any unattended tests to safeguard samples. - Clean Interior Surfaces
Clean the mirror-finish stainless-steel interior surfaces with appropriate non-abrasive cleaners to maintain hygiene and performance.
What is the effective temperature range of the LHS-50CH humidity chamber without refrigeration?
The LHS-50CH (CH configuration) operates from RT+10°C to 85°C, with no refrigeration function. This means the minimum achievable temperature is 10°C above the ambient room temperature, typically limiting low-temperature testing to around 25-40°C depending on lab conditions.
Can the chamber control humidity below 80% RH?
The humidity control range is specified as 80%RH to 95%RH, with a deviation of +/-3%RH. The chamber cannot control humidity below 80%RH, as it uses a balanced control system without dehumidification capability. For lower humidity requirements, consider refrigerated models (CL, CA, CB) that may offer broader humidity ranges.
What are the space and power requirements for installing the LHS-50CH chamber?
The chamber external dimensions are 720 x 620 x 725 mm, requiring a clear installation footprint. It requires a single-phase AC 220V, 50Hz power supply with a rated input of 1450W. The working environment must be maintained at +5 to 30°C. Ensure adequate ventilation and access to the side test port, drain, and controller surfaces for maintenance.
The LHS-50CH 50 L balanced-control constant temperature and humidity chamber provides stable RT+10 to 85°C and 80–95%RH control with mirror-finish stainless steel interior, but lacks refrigeration (cannot cool below ambient) and cannot achieve humidity below 80%RH.
Positive
- Balancing temp-humidity control: Microcomputer balancing controller maintains stable operating conditions for incubation, cultivation, and stability studies, reducing drift during long-duration tests.
- Mirror-finish stainless steel interior: Smooth, corrosion-resistant inner chamber with removable shelving simplifies cleaning and decontamination between test cycles, supporting laboratory hygiene and repeatability.
Trade-offs
- No refrigeration function: This CH model cannot cool below room temperature (minimum setpoint RT+10°C), limiting its use to above-ambient conditions and excluding low-temperature testing.
- Restricted humidity range: Humidity control is specified only from 80%RH to 95%RH, preventing low-humidity or desiccating experiments and requiring alternative equipment for drier conditions.