Programmable Climate Chamber, 360 L
Research-grade laboratory product
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||
|
LABORATORY PROCUREMENT SUPPORT
For model selection, accessory matching, installation-space review, power-supply confirmation or configuration support, contact our technical sales team.
E-MAIL: inquiry@atomfair.com
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Supplier: Atomfair
Brand: ATOMFAIR®
|
The chamber must be operated within specified ambient temperature limits to maintain performance. When set temperature is 10°C or below, ambient temperature should not exceed 26°C, and when setpoint is above 40°C, humidity setpoints should generally not exceed 80%RH.
- Ambient Temperature Constraint: When the set temperature is 10°C or below, the ambient temperature must not exceed 26°C to ensure proper operation.
- Humidity Setpoint Constraint: When the temperature setpoint exceeds 40°C, the humidity setpoint should generally not exceed 80%RH to avoid performance issues.
- Power Supply Requirement: The chamber requires a grounded AC220V 50Hz power supply with a 900 W input capacity.
- Transport and Positioning: Keep the unit upright during unloading and positioning, and maintain stable room temperature away from direct heat sources.
Position the chamber in a stable environment away from direct heat sources and connect to the appropriate power supply. Set the programmable parameters and allow the chamber to stabilize before introducing samples.
Required Equipment: LCD microcomputer programmable controller, Chamber body
- Position the chamber
Position the chamber upright on a level surface away from direct heat sources and ensure adequate ventilation. - Connect to power supply
Connect the chamber to a grounded AC220V 50Hz power supply and verify that the ambient room temperature is stable and within the required range. - Set control parameters
Set the desired temperature, humidity, and light intensity parameters using the LCD microcomputer programmable controller. - Allow chamber stabilization
Allow the chamber to reach the setpoint and stabilize for at least 30 minutes before placing samples inside. - Monitor chamber conditions
Monitor chamber conditions periodically and adjust parameters as needed to maintain the required experimental environment.
What are the temperature and humidity performance limits of the Atomfair AF-QHX360T programmable climate chamber?
The AF-QHX360T provides a no-light temperature range of 5 to 60 °C and an illuminated range of 10 to 60 °C, with 0.1 °C resolution and ±2 °C uniformity. Humidity is controllable from 50%RH to 90%RH with ±8%RH deviation. These figures are based on no-load testing at 20 °C and 50%RH ambient humidity.
Can the chamber maintain high humidity when the temperature setpoint is above 40 °C?
The chamber is not specified for high-humidity operation above 40 °C; the configuration note states that when the temperature setpoint is above 40 °C, humidity setpoints should generally not exceed 80%RH. Additionally, when the set temperature is 10 °C or below, ambient temperature should not exceed 26 °C.
What power supply and installation conditions are required for the 360 L programmable climate chamber?
The chamber requires AC220V 50Hz power with a 900 W input rating. It should be kept away from direct heat sources in a stable room-temperature environment, and the unit must remain upright during unloading and positioning. The external footprint is 835 x 755 x 1695 mm and net weight is 129.5 kg, so floor loading and access space should be confirmed before installation.
The Atomfair AF-QHX360T is a 360 L programmable climate chamber with top lighting, temperature range 5-60°C (no-light) or 10-60°C (illuminated), humidity 50-90%RH, and light up to 10000 lux. Its stainless-steel interior and programmable controller support plant growth and biological studies. However, ambient temperature must be ≤26°C when set temperature is ≤10°C, and humidity should not exceed 80% at setpoints above 40°C.
Positive
- Programmable multi-parameter control: The LCD microcomputer controller enables programmable temperature, light, humidity, and time settings, allowing users to simulate natural climate conditions for cultivation, germination, and tissue culture experiments.
- Top lighting for uniform illumination: Top-mounted lighting with 6 intensity levels up to 10000 lux supports repeatable photoperiod experiments and even light distribution across the 360 L chamber, beneficial for plant growth and seed germination studies.
Trade-offs
- ambient temperature constraint at low setpoint: When the chamber is set to 10°C or below, the ambient laboratory temperature must not exceed 26°C to maintain performance, which may require additional climate control in warm facilities.
- Humidity limitation at high temperature: At set temperatures above 40°C, the humidity setpoint should generally not exceed 80%RH, restricting simultaneous high-temperature and high-humidity operation for certain protocols.
Every advanced material, component, equipment, and instrument in our catalog is backed by rigorous testing. We maintain strict internal quality management frameworks and align with CE conformity metrics to deliver transparent, reproducible performance data via our public open-science repository.
To request raw batch performance data, submit formal vendor registration paperwork, or execute a fast-turnaround R&D manufacturing loop, contact us at inquiry@atomfair.com.
Item is dispatched under the Atomfair Shipping & Delivery Framework (Free worldwide shipping on orders over $59 USD excl. heavy equipment). Return is governed by the Atomfair Return & Refund Policy (7-day technical return window).