Programmable Artificial Climate Chamber, 1500 L
Research-grade laboratory product
|
|||||||||||||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||||||||||||
|
LABORATORY PROCUREMENT SUPPORT
For model selection, accessory matching, installation-space review, power-supply confirmation or configuration support, contact our technical sales team.
E-MAIL: inquiry@atomfair.com
|
|||||||||||||||||||||||||||||||||||||||||||||||||
|
Supplier: Atomfair
Brand: ATOMFAIR®
|
Operate the chamber within an ambient temperature range of +5 to 30°C and supply AC 220V, 50 Hz power with adequate capacity for 4500 W. Keep the chamber upright during transport and positioning, and maintain clear access to air inlets, exhaust paths, and control interfaces for proper operation.
- Electrical Supply: Connect the chamber to a dedicated AC 220V, 50 Hz power supply rated for at least 4500 W to avoid circuit overload.
- Ambient Temperature: Operate the chamber only when the ambient temperature is between +5°C and 30°C to ensure proper cooling system performance.
- Transport Orientation: Transport and store the chamber in an upright orientation to prevent compressor oil migration and damage.
- Ventilation: Keep the air inlets and exhaust paths unobstructed to maintain proper airflow and prevent overheating.
Install the chamber in a suitable location with adequate space and power supply, then configure the controller for the desired environmental parameters. Follow these steps to ensure safe setup and reliable operation.
Required Equipment: Lifting equipment
- Verify Installation Space
Confirm that the installation space accommodates the external dimensions of 1570 x 1475 x 2050 mm. - Prepare Power Supply
Prepare a dedicated AC 220V, 50 Hz power supply capable of supporting the 4500 W load. - Position the Chamber
Unpack the chamber while keeping it upright and use lifting equipment to move it into the installation space. - Ensure Accessibility
Ensure that air inlets, exhaust paths, shelves, ports, doors, and control interfaces remain accessible for cleaning and maintenance. - Configure Controller
Program the multi-segment controller with the desired temperature, humidity, light intensity, and time settings. - Connect Optional Accessories
Connect optional communication interfaces, CO2 inlet, or alarm system as required for the application. - Perform System Check
Verify that all parameters are set correctly and that the chamber operates within specified tolerances before loading samples.
What is the actual temperature range of the 1500 L artificial climate chamber when lighting is active compared to when it is off?
With lighting active, the temperature range is 10 to 50 °C; without lighting, the range extends from 0 to 50 °C. Temperature resolution is 0.1 °C with a fluctuation/uniformity of ±1 °C. These ranges are explicitly specified for the 1500 L model and reflect the impact of lighting heat on chamber performance.
Can the chamber be integrated with external data logging or alarm systems for remote monitoring?
Yes, the 1500 L chamber supports optional RS485 or 232 communication interface and software, USB data storage, an embedded printer, and a wireless alarm system. These options enable remote data logging, trend review, and immediate fault notification without requiring aftermarket modifications.
What are the critical installation space and power requirements for this 1500 L chamber?
Installation requires a minimum external footprint of 1570 x 1475 x 2050 mm and a dedicated 4500 W power supply at AC 220 V, 50 Hz. The working environment must be maintained between +5 and 30 °C. Keeping air inlets, exhaust paths, and doors accessible is essential for proper operation and maintenance.
This 1500 L programmable climate chamber features dual alternating compressors for long-duration continuous operation and multi-segment control of temperature, humidity, and light. The 4500 W power requirement and +/-5-7%RH humidity deviation are practical constraints for lab infrastructure planning.
Positive
- Dual alternating compressor design: Two imported compressors automatically alternate to support continuous long-duration operation, reducing wear and improving reliability for extended cultivation runs.
- Multi-segment programmable control: Independent setpoint programming for temperature, humidity, light intensity, and time in up to 30 segments allows precise replication of climate profiles for plant growth, tissue culture, and stability studies.
Trade-offs
- High power consumption: The 4500 W input at 220 V/50 Hz requires dedicated high-current electrical infrastructure, which may not be available in all laboratory settings.
- Humidity deviation tolerance: Specified humidity deviation of +/-5 to 7%RH limits precision for applications requiring tight humidity control, such as certain seed germination or tissue culture protocols.
Every advanced material, component, equipment, and instrument in our catalog is backed by rigorous testing. We maintain strict internal quality management frameworks and align with CE conformity metrics to deliver transparent, reproducible performance data via our public open-science repository.
To request raw batch performance data, submit formal vendor registration paperwork, or execute a fast-turnaround R&D manufacturing loop, contact us at inquiry@atomfair.com.
Item is dispatched under the Atomfair Shipping & Delivery Framework (Free worldwide shipping on orders over $59 USD excl. heavy equipment). Return is governed by the Atomfair Return & Refund Policy (7-day technical return window).