LED Plant Growth Chamber, 350 L, with Humidity Control
Research-grade laboratory product
|
|||||||||||||||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||||||||||||||
|
LABORATORY PROCUREMENT SUPPORT
For model selection, accessory matching, installation-space review, power-supply confirmation or configuration support, contact our technical sales team.
E-MAIL: inquiry@atomfair.com
|
|||||||||||||||||||||||||||||||||||||||||||||||||||
|
Supplier: Atomfair
Brand: ATOMFAIR®
|
Operate the chamber within the specified temperature and humidity ranges to avoid equipment damage and ensure safe operation. Maintain the working environment between 0 and 35 °C and provide a stable AC 220 V, 50 Hz power supply with adequate capacity.
- Installation Space: Ensure the chamber is placed with the external dimensions 810 × 835 × 1840 mm and keep air inlets, exhaust paths, shelves, ports, doors, and control interfaces accessible for ventilation and maintenance.
- Transport Handling: Keep the packed chamber upright during transport and use suitable lifting equipment to avoid impact damage.
- Power Supply: Connect the chamber only to an AC 220 V, 50 Hz power supply rated for at least 2000 W to prevent electrical overload.
- Ambient Conditions: Operate the chamber within an ambient temperature range of 0 to 35 °C to maintain control stability and prevent compressor stress.
- Humidity Range: Set humidity within the range of 50% to 90% RH as specified to maintain control accuracy and avoid condensation.
Follow these steps to safely install, configure, and begin operating the plant growth chamber. Proper setup ensures stable environmental conditions and prolongs equipment life.
Required Equipment: suitable lifting equipment
- Confirm Installation Space and Power
Confirm that the installation location meets the external dimensions of 810 × 835 × 1840 mm and that a dedicated AC 220 V, 50 Hz power supply is available. - Unpack and Position Chamber
Unpack the chamber while keeping it upright and use suitable lifting equipment to move it into position with accessible inlets, exhaust paths, and control interfaces. - Connect Power and Initialize
Connect the chamber to the power supply, switch on the main power, and verify that the 7-inch touchscreen controller powers on and displays the default menu. - Configure Temperature Setpoint
Set the temperature according to the installed light decks: without light use 5–45 °C, with single-layer light use 10–45 °C, or with double-layer light use 15–45 °C. - Set Humidity Target
Set the humidity target between 50% and 90% RH and confirm that the controller stabilizes within the specified deviation of ±3% RH at 80% RH. - Load Samples and Begin Operation
Load samples onto the adjustable shelves without blocking airflow, close the door securely, and start the programmed cycle or manual control.
How does the temperature control range change when using different lighting configurations in the Atomfair AF-PGH0350 LED Plant Growth Chamber?
The temperature control range depends on the lighting configuration: without light it is 5 to 45°C, with single-layer light it is 10 to 45°C, and with double-layer light it is 15 to 45°C. This trade-off means that using more light decks raises the minimum achievable temperature, which may affect low-temperature experiments requiring high illumination.
What is the minimum achievable humidity in the Atomfair AF-PGH0350 chamber?
The chamber controls humidity from 50%RH to 90%RH. It cannot achieve humidity levels below 50%RH, so applications requiring low humidity (e.g., desiccation studies) are not supported without additional equipment.
What are the electrical and space requirements for installing the Atomfair AF-PGH0350 LED Plant Growth Chamber?
The chamber requires a 2000 W power supply at AC 220 V, 50 Hz, and external dimensions of 810 x 835 x 1840 mm. Ensure adequate ventilation and access for maintenance; the unit is shipped upright in reinforced packaging and requires suitable lifting equipment.
The LED Plant Growth Chamber 350L with humidity control delivers dual-imported compressor reliability, SUS304 stainless steel interior, and programmable multi-segment control, but temperature range is constrained by lighting load and humidity is limited to 50–90%RH.
Positive
- Dual imported compressors for long-duration stability: The chamber uses dual imported compressors with a long-duration operation design, supporting dependable low-temperature performance and sustained running stability for continuous cultivation or climate simulation.
- SUS304 stainless steel chamber interior: The SUS304 stainless-steel interior and adjustable shelf supports provide corrosion resistance and easy cleaning, supporting reliable day-to-day use in plant growth, tissue culture, and seed germination workflows.
Trade-offs
- Temperature range constrained by lighting load: The temperature control range shifts from 5–45°C without light to 15–45°C with double-layer light, limiting low-temperature operation when full illumination is required.
- Humidity range limited to 50–90%RH: The humidity function covers only 50%RH to 90%RH, making the chamber unsuitable for experiments requiring low-humidity conditions below 50%RH.
Every advanced material, component, equipment, and instrument in our catalog is backed by rigorous testing. We maintain strict internal quality management frameworks and align with CE conformity metrics to deliver transparent, reproducible performance data via our public open-science repository.
To request raw batch performance data, submit formal vendor registration paperwork, or execute a fast-turnaround R&D manufacturing loop, contact us at inquiry@atomfair.com.
Item is dispatched under the Atomfair Shipping & Delivery Framework (Free worldwide shipping on orders over $59 USD excl. heavy equipment). Return is governed by the Atomfair Return & Refund Policy (7-day technical return window).