Temperature and Humidity Chamber, 247 L
Research-grade laboratory product
|
|||||||||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||||||||
|
LABORATORY PROCUREMENT SUPPORT
For model selection, accessory matching, installation-space review, power-supply confirmation or configuration support, contact our technical sales team.
E-MAIL: inquiry@atomfair.com
|
|||||||||||||||||||||||||||||||||||||||||||||
|
Supplier: Atomfair
Brand: ATOMFAIR®
|
When the set temperature is 10°C or below, ambient temperature must not exceed 26°C. When the set temperature is above 40°C, the humidity setpoint should generally not exceed 80% to protect service life.
- Temperature-Dependent Ambient Limit: At setpoints at or below 10°C, ambient room temperature must not exceed 26°C to maintain chamber stability.
- High-Temperature Humidity Constraint: When the chamber is operating above 40°C, the humidity setpoint must be kept at or below 80% to prevent damage to the system.
Position the chamber upright with clear air paths and sufficient service space for ventilation and maintenance. Program temperature, humidity, and time parameters using the LCD controller for repeatable workflows.
Required Equipment: Temperature and Humidity Chamber (247 L, 5 to 50°C)
- Position the Chamber
Place the unit upright on a level surface with at least 15 cm of clearance on all sides for airflow and maintenance access. - Connect Power Supply
Connect the chamber to a grounded AC 220 V, 50 Hz outlet rated for 800 W continuous load. - Set Temperature Conditions
Enter the desired temperature setpoint (5 to 50°C) via the LCD controller, ensuring ambient temperature does not exceed 26°C if setpoint is 10°C or below. - Configure Humidity Parameters
Program the humidity setpoint (50% to 90% RH) on the LCD, keeping it at or below 80% if the temperature is above 40°C. - Verify Chamber Stability
Allow the chamber to stabilize for 30 minutes and confirm the LCD display shows steady temperature and humidity within specified fluctuation limits.
What are the operational constraints for temperature and humidity setpoints in the Atomfair AF-HSP250BE 247 L chamber?
The chamber’s humidity control range is 50%RH to 90%RH, but when the set temperature exceeds 40 °C, the humidity setpoint should generally not exceed 80%RH to protect equipment service life. Additionally, when the set temperature is 10 °C or below, the ambient room temperature must not exceed 26 °C. These constraints are explicitly stated in the technical specifications and important notice.
Is the Atomfair AF-HSP250BE suitable for BOD measurement and microbial culture?
Yes, the product description lists BOD measurement and microbial culture as intended applications, along with environmental protection, veterinary research, drug testing, aquatic research, plant cultivation, and breeding tests. The 247 L capacity, 304 stainless steel interior, programmable LCD controller, and ducted air circulation system support these workflows by providing steady, repeatable temperature and humidity conditions.
What power and installation requirements are needed for the Atomfair AF-HSP250BE humidity chamber?
The chamber requires an AC 220 V, 50 Hz power supply with an input of 800 W. Its external dimensions are 740 × 705 × 1560 mm and net weight is 106.5 kg. The unit must be kept upright during handling and positioning, with clear air paths and sufficient service space around the unit for ventilation and maintenance. Shipping uses reinforced packaging, and freight is quoted separately.
The Temperature and Humidity Chamber, 247 L (AF-HSP250BE) provides 247 L of usable space with temperature control from 5 to 50°C and humidity from 50% to 90%RH, featuring a 304 stainless steel chamber, self-diagnosis, and power-off memory. However, operation below 10°C requires ambient temperature ≤26°C, and when set above 40°C, humidity setpoints must be kept below 80% to protect service life.
Positive
- Self-diagnosis and over-temperature protection: Built-in self-diagnosis and over-temperature self-limiting functions help protect samples and reduce operational risk, supporting reliable day-to-day chamber use.
- Power-off memory for continuous workflows: Power-off memory preserves temperature, humidity, and time settings after interruptions, allowing uninterrupted operation and repeatable experimental conditions.
Trade-offs
- Low-temperature ambient constraint: When set temperature is 10°C or below, ambient temperature must not exceed 26°C, limiting placement in unconditioned spaces or hot labs.
- High-temperature humidity limitation: When set temperature is above 40°C, humidity setpoints should generally not exceed 80% to protect service life, restricting combined high-temperature/high-humidity tests.