240 L Water-Jacket UV plus 90 °C Sterilization CO2 Incubator, High-End Touchscreen Series
Research-grade laboratory product
|
|||||||||||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||||||||||
|
LABORATORY PROCUREMENT SUPPORT
For model selection, accessory matching, installation-space review, power-supply confirmation or configuration support, contact our technical sales team.
E-MAIL: inquiry@atomfair.com
|
|||||||||||||||||||||||||||||||||||||||||||||||
|
Supplier: Atomfair
Brand: ATOMFAIR®
|
Operate the incubator within an ambient temperature range of +5 to +30°C with a dedicated AC 220V 50 Hz power supply. The chamber must remain upright during transport and installation, with all serviceable components accessible for routine cleaning and inspection.
- Accessibility Constraint: Ensure that all serviceable components including sensors, shelves, filters, doors, test ports, and control interfaces remain accessible for cleaning and inspection.
- Power Supply Constraint: Use only the specified AC 220V 50 Hz power supply and avoid overloading the circuit.
- Transport Constraint: Keep the packed unit upright during transport and use suitable lifting equipment to prevent impact damage.
- Accessory Selection Constraint: Select optional accessories such as O2 control or humidity display only after confirming compatibility with the workflow.
- Alarm Monitoring Constraint: Monitor the alarm system for power failure, CO2 deviation, door opening, and temperature sensor faults to mitigate contamination risks.
Position the incubator in a well-ventilated area with stable ambient temperature and a dedicated power supply. After installation, program the desired temperature, CO2 level, and sterilization cycle via the touchscreen interface before introducing cultures.
Required Equipment: Lifting equipment, Touchscreen controller
- Confirm installation space and power supply
Confirm that the installation space accommodates the incubator's external dimensions of 754×760×1020 mm and that a dedicated AC 220V 50 Hz power outlet is available. - Inspect the incubator
Inspect the chamber, sensors, shelves, filters, and doors for any signs of damage or contamination before powering on the unit. - Connect power supply
Connect the incubator to the power supply and verify that the touchscreen display activates and shows the home screen. - Set temperature and CO2
Set the target temperature within the range of RT+3 to 55°C and the CO2 concentration between 0% and 20% using the touchscreen controller. - Run sterilization cycle
Select the appropriate sterilization mode (UV or 90°C high-temperature) via the touchscreen and run the sterilization cycle before introducing any cultures.
What is the CO2 recovery time for this 240 L water-jacket CO2 incubator after a door opening?
The CO2 recovery time is no more than 3 minutes to return to 5% CO2 after a 30-second door opening, as specified in the technical specifications. This rapid recovery minimizes disruption to the culture environment, supporting stable cell growth conditions.
What power supply and ambient temperature conditions are required for installing this incubator?
The incubator requires AC 220 V, 50 Hz power supply and an ambient temperature range of +5 to +30 °C. The external dimensions are 754 × 760 × 1020 mm, so sufficient space and ventilation must be confirmed before installation.
What sterilization and alarm features are included in the UV plus 90 °C sterilization model?
This model (RWP(90)) includes both UV and 90 °C high-temperature high-humidity sterilization for chamber hygiene. The alarm system covers power failure, CO2 deviation, door opening, and multiple temperature sensor conditions, with an optional wireless alarm for unattended monitoring.
The 240 L water-jacket CO2 incubator with UV and 90°C sterilization provides ±0.1°C temperature stability and ≤3 min CO2 recovery to 5% after door opening, with a touchscreen interface for traceable monitoring. Operation requires AC 220V/50Hz power and ambient temperatures between 5–30°C; humidity is via natural evaporation with optional display only.
Positive
- Infrared CO2 sensing with fast recovery: Infrared CO2 sensor provides ±0.1% accuracy and CO2 recovery to 5% within 3 minutes after a 30-second door opening, maintaining stable culture conditions.
- Touchscreen with trend and alarm records: Large touchscreen interface supports curve display, alarm review, and 72-hour running record lookup for real-time parameter monitoring and traceability.
Trade-offs
- Restricted ambient temperature range: Working ambient temperature must be between +5°C and +30°C; operation outside this range may affect temperature control stability.
- Natural evaporation humidity; no active control: Relative humidity is achieved via natural evaporation (≥90% typical) with only optional display; no active humidity control is available.
Every advanced material, component, equipment, and instrument in our catalog is backed by rigorous testing. We maintain strict internal quality management frameworks and align with CE conformity metrics to deliver transparent, reproducible performance data via our public open-science repository.
To request raw batch performance data, submit formal vendor registration paperwork, or execute a fast-turnaround R&D manufacturing loop, contact us at inquiry@atomfair.com.
Item is dispatched under the Atomfair Shipping & Delivery Framework (Free worldwide shipping on orders over $59 USD excl. heavy equipment). Return is governed by the Atomfair Return & Refund Policy (7-day technical return window).