HiPette-LTS Single-channel 5-50 µL PipetteProduct Type: Fully autoclavable manual pipette
Research-grade liquid handling platform
|
|||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||
|
TAILORED SOLUTIONS FOR RESEARCH
For bulk orders, OEM requirements, accessory matching, platform selection or configuration confirmation, contact our engineering sales team.
EMAIL: inquiry@atomfair.com
|
|||||||||||||||||||||||||||||||
|
Supplier: Atomfair
Brand: ATOMFAIR®
|
The pipette must be calibrated according to ISO 8655-2 to maintain specified accuracy. It is fully autoclavable for sterilization between uses.
- Calibration Requirements: Calibration must be performed at the three test volumes (50, 25, and 5 µL) to verify that systematic and random error remain within the specified limits.
- Tip Compatibility: Only compatible pipette tips should be used to ensure proper sealing and accuracy; incompatible tips may cause leakage or inaccurate volume delivery.
How does the systematic error of the HiPette-LTS 5-50 µL pipette vary across its volume range, and what is the trade-off between accuracy at high versus low volumes?
Systematic error increases significantly at lower volumes: at 50 µL it is ±0.45 µL (±0.9%), at 25 µL ±0.35 µL (±1.4%), and at 5 µL ±0.15 µL (±3.0%). This means the percentage error is three times higher at 5 µL than at 50 µL, a trade-off that should be considered when planning assays requiring high precision near the lower end of the range.
What calibration standard does the HiPette-LTS 5-50 µL pipette adhere to, and what are the implications for integration into ISO-compliant laboratory workflows?
The pipette includes a user calibration reference per ISO 8655-2, as indicated in the technical table. This ensures compatibility with laboratories that require traceable calibration documentation for method validation, but users must perform their own calibration checks using the referenced standard to maintain ISO compliance.
What are the autoclaving requirements for the HiPette-LTS 5-50 µL pipette, and are there any restrictions on the number of sterilization cycles?
The pipette is described as fully autoclavable, but the product description does not specify maximum cycle count or exact autoclave parameters. Standard laboratory practice for fully autoclavable pipettes is typically 121°C for 20 minutes; for validated cycle limits and long-term performance, contact Atomfair technical support.
The HiPette-LTS single-channel 5-50 µL pipette provides full autoclavability and good repeatability at higher volumes, but its systematic error increases substantially at the lower end of the range, making it best suited for mid-to-high volume transfers where precision is less critical.
Positive
- Fully autoclavable design: The entire pipette can be autoclaved at 121°C, enabling thorough sterilization between uses to prevent cross-contamination in sensitive workflows.
- Low random error at full volume: At 50 µL, random error is ±0.15 µL (±0.3%), ensuring excellent repeatability for routine transfer and assay setup.
Trade-offs
- High systematic error at low volume: At 5 µL, systematic error reaches ±0.15 µL (±3.0%), significantly reducing accuracy for low-volume applications and requiring careful volume selection.
- Calibration and tip dependency: The user must verify calibration per ISO 8655-2 and ensure compatibility with selected tips to maintain specified performance, adding pre-use validation steps.