Description
SpectraCore UV 2101 UV-Visible Spectrophotometer, 190-1100 nm, 1.0 nmProduct Type: UV-Visible Spectrophotometer
Research-grade laboratory product
|
|||||||||||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||||||||||
|
LABORATORY PROCUREMENT SUPPORT
For model selection, accessory matching, platform compatibility or configuration confirmation, contact our technical sales team.
E-MAIL: inquiry@atomfair.com
|
|||||||||||||||||||||||||||||||||||||||||||||||
|
Manufacturer: Atomfair LLC
Brand: ATOMFAIR®
|
The SpectraCore UV 2101 requires a 110/220 V, 50/60 Hz, 120 W power supply for operation. The instrument uses a single-beam optical system with a 4-position cell holder for 10 mm pathlength cuvettes, with a fixed long-cell holder option available.
- Power Supply: Connect the instrument to a grounded power outlet providing 110/220 V, 50/60 Hz, 120 W; do not use with ungrounded or non-specification power sources.
- Sample Holder Compatibility: The standard sample holder accepts only 10 mm pathlength cuvettes in a 4-position cell holder; alternative cuvette formats require the optional fixed long-cell holder.
How does the photometric repeatability of the SpectraCore UV 2101 vary across the absorbance range?
The photometric repeatability is specified as ±0.001 A from 0 to 0.5 A and ±0.0015 A from 0.5 to 1 A, indicating higher precision at lower absorbance values. This specification supports reliable quantitative analysis within the linear absorbance range.
What cuvette path lengths are compatible with the SpectraCore UV 2101 sample holder?
The standard 4-position cell holder accepts 10 mm path length cuvettes. A fixed long-cell holder option is available for measurements requiring longer path lengths, providing flexibility for analyzing dilute samples.
What are the bench space and power requirements for installing the SpectraCore UV 2101?
The instrument measures 570 x 440 x 210 mm and weighs 20 kg, requiring a stable laboratory bench. It operates on 110/220 V, 50/60 Hz with a power consumption of 120 W.
The SpectraCore UV 2101 is a research-grade single-beam spectrophotometer offering 1.0 nm bandwidth and high wavelength accuracy across the UV-Vis range, suitable for standard absorbance and transmittance analyses. Its low stray light and stable photometric performance support reliable quantitative measurements.
Positive
- High wavelength accuracy and repeatability: Wavelength accuracy of ±0.1 nm at 656.1 nm and ±0.3 nm full range, with repeatability ≤0.1 nm, ensures precise spectral positioning for quantitative analysis.
- Low stray light and stable photometric performance: Stray light ≤0.03%T at 220 and 360 nm (≤0.02%T on selected models) and photometric stability ≤0.001 A/h at 500 nm enable reliable absorbance measurements even at high optical densities.
Trade-offs
- Single-beam optical system requires baseline correction: The single-beam design necessitates periodic baseline correction before each measurement set to compensate for lamp drift and environmental changes, adding procedural steps.
- Stray light may limit accuracy at high absorbance: With stray light ≤0.03%T at key wavelengths, accuracy at absorbance values above 2 A may be degraded compared to instruments with lower stray light specifications.
Every advanced material, component, equipment, and instrument in our catalog is backed by rigorous testing. We maintain strict internal quality management frameworks and align with CE conformity metrics to deliver transparent, reproducible performance data via our public open-science repository.
To request raw batch performance data, submit formal vendor registration paperwork, or execute a fast-turnaround R&D manufacturing loop, contact us at inquiry@atomfair.com.
Item is dispatched under the Atomfair Shipping & Delivery Framework (Free worldwide shipping on orders over $59 USD excl. heavy equipment). Return is governed by the Atomfair Return & Refund Policy (7-day technical return window).