MicroPette Plus Single-channel 50-200 µL PipetteProduct Type: Fully autoclavable manual pipette
Research-grade liquid handling platform
|
|||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||
|
TAILORED SOLUTIONS FOR RESEARCH
For bulk orders, OEM requirements, accessory matching, platform selection or configuration confirmation, contact our engineering sales team.
EMAIL: inquiry@atomfair.com
|
|||||||||||||||||||||||||||||||
|
Supplier: Atomfair
Brand: ATOMFAIR®
|
This pipette requires periodic calibration per ISO 8655-2 to maintain specified accuracy. Autoclaving at standard conditions is required for sterilization; ensure complete drying before use.
- Calibration Requirement: Calibrate the pipette at 200 µL, 100 µL, and 50 µL test volumes using a gravimetric method per ISO 8655-2.
- Tip Compatibility: Use only tips designed for the 50-200 µL volume range to maintain proper seal and accuracy.
- Sterilization Procedure: Autoclave the fully disassembled pipette at 121°C for 20 minutes, then allow to dry completely before reassembly.
- Volume Setting Constraint: Set volume by rotating the plunger button; do not exceed the 200 µL maximum or go below the 50 µL minimum.
This procedure describes the correct technique for liquid handling using the MicroPette Plus pipette. Follow these steps to ensure accurate and reproducible volume delivery.
Required Equipment: Pipette tips compatible with 50-200 µL range
- Set Volume
Set the volume by rotating the plunger button to the desired value within the 50-200 µL range. - Attach Tip
Attach a compatible tip by pressing the pipette shaft firmly into the tip. - Aspirate Liquid
Aspirate liquid by depressing the plunger to the first stop, immersing the tip, and slowly releasing the plunger. - Dispense Liquid
Dispense liquid by depressing the plunger to the first stop, then to the second stop to expel any residual liquid. - Eject Tip
Eject the used tip by pressing the tip ejector button.
How does the systematic and random error of the MicroPette Plus 50-200 µL pipette change across the working volume range?
At 200 µL, systematic error is ±0.60% and random error ±0.15%. At 100 µL, systematic error is ±0.80% and random error ±0.30%. At 50 µL, systematic error is ±1.00% and random error ±0.40%. Relative error increases as set volume decreases, so for highest accuracy, operate near the upper end of the range.
Is the MicroPette Plus 50-200 µL pipette calibrated to ISO 8655-2 and can it be user-calibrated?
Yes, the pipette is calibrated with reference to ISO 8655-2, and the technical table indicates a user calibration reference. This allows users to perform calibration checks and adjustments in-house using standard gravimetric methods.
Can the MicroPette Plus 50-200 µL pipette be sterilized by autoclaving?
Yes, the pipette is fully autoclavable, allowing complete sterilization without disassembly. This is explicitly stated in the product specifications under sterilization suitability.
The MicroPette Plus Single-channel 50-200 µL pipette provides fully autoclavable construction and ISO 8655-2 calibration traceability, but exhibits reduced accuracy at the 50 µL lower limit and requires compatible tips and calibration maintenance for reliable operation.
Positive
- Fully autoclavable without disassembly: The entire pipette can be sterilized by autoclaving at 121°C, eliminating the need for disassembly and reducing contamination risk in aseptic workflows.
- ISO 8655-2 calibration traceability: Calibration reference to ISO 8655-2 provides documented accuracy and precision data for method validation and quality assurance.
Trade-offs
- Accuracy degrades at low volume: Systematic error at 50 µL is ±1.00% compared to ±0.60% at 200 µL, requiring careful volume selection for critical assays.
- Requires compatible tips and calibration: Optimal performance depends on matched tips and periodic calibration verification, adding operational overhead and consumable compatibility checks.
Every advanced material, component, equipment, and instrument in our catalog is backed by rigorous testing. We maintain strict internal quality management frameworks and align with CE conformity metrics to deliver transparent, reproducible performance data via our public open-science repository.
To request raw batch performance data, submit formal vendor registration paperwork, or execute a fast-turnaround R&D manufacturing loop, contact us at inquiry@atomfair.com.
Item is dispatched under the Atomfair Shipping & Delivery Framework (Free worldwide shipping on orders over $59 USD). Return is governed by the Atomfair Return & Refund Policy (7-day technical return window).