Description
IoniSense Conductivity 100 T Touch Screen Benchtop Electrochemistry MeterProduct Type: Touch Screen Benchtop Electrochemistry Meter
Research-grade laboratory product
|
|||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||
|
LABORATORY PROCUREMENT SUPPORT
For model selection, accessory matching, platform compatibility or configuration confirmation, contact our technical sales team.
E-MAIL: inquiry@atomfair.com
|
|||||||||||||||||||||||||||||||||||||
|
Manufacturer: Atomfair LLC
Brand: ATOMFAIR®
|
This instrument requires calibration at one point using certified conductivity standard solutions. Temperature compensation is automatically applied within the range 0.0-50.0 °C.
- Calibration requirement: The meter must be calibrated with a single standard solution before each measurement session to ensure accuracy.
- Electrode compatibility: Only the specified IoniProbe Conductivity 201 TG or 201 T electrode or an equivalent with matching connector and temperature sensor should be used.
- Power supply constraints: The instrument is powered by a DC12V adapter and must not be used with alternate power sources.
- Measurement range autosegmentation: The conductivity range is automatically segmented into five sub-ranges to optimize resolution across 0-200.0 mS/cm.
- Data output interface: Measurement data can be exported via USB to a PC or printer for documentation.
This procedure describes the calibration and measurement workflow for the IoniSense Conductivity 100 T. Follow these steps to obtain accurate readings.
Required Equipment: IoniSense Conductivity 100 T meter, IoniProbe conductivity electrode, Conductivity standard solution, Magnetic stirrer
- Connect electrode
Attach the IoniProbe conductivity electrode to the meter's input port and secure the connection. - Calibrate
Immerse the electrode in the conductivity standard solution, stir gently, and initiate the one-point calibration routine. - Measure sample
Rinse the electrode with deionized water, immerse it in the sample, and record the displayed conductivity value after stabilization. - Export data
Connect the meter to a PC or printer via USB to transfer logged measurements for further analysis.
How does the single-point calibration affect accuracy across the conductivity meter's automatic segmentation ranges?
The IoniSense Conductivity 100 T uses a single-point calibration (1 point) for its entire conductivity range of 0–200.0 mS/cm, which is divided into six automatic segments (e.g., 0–19.99 µS/cm, 20.0–199.9 µS/cm, etc.). The stated accuracy is ±0.5% FS ±1 digit, so calibration at a single point may not compensate for non-linearities across extremely different measurement ranges. For best results, calibrate at a conductivity value representative of the expected sample. No multi-point calibration option is indicated in the source.
Which electrodes are compatible with the Conductivity 100 T meter, and what are the limitations for temperature compensation?
The standard package includes either the IoniProbe Conductivity 201 TG or IoniProbe Conductivity 201 T conductivity electrode. The meter's temperature compensation range is (0.0–50.0) °C / (32.0–122.0) °F, and the electrode's own temperature rating may differ. The source does not specify the electrode's cell constant or maximum temperature, so users must confirm the electrode matches the application and operating temperature before purchase.
What are the power and accessory requirements for setting up the IoniSense Conductivity 100 T in a laboratory?
The meter requires a DC12V power supply, which is included in the standard package. The package also contains the instrument, a conductivity electrode, conductivity standard solutions, a universal electrode holder, and a magnetic stirrer. No specific storage conditions are provided in the source, but as a benchtop electrochemistry meter, it should be kept in a clean, dry environment. Ensure the laboratory has a standard DC12V outlet or use the included adapter.
The IoniSense Conductivity 100 T is a benchtop electrochemistry meter supporting conductivity, TDS, salinity, resistivity, and temperature measurements with automatic range segmentation up to 200 mS/cm, suitable for water quality screening and laboratory verification workflows.
Positive
- Multi-parameter measurement capability: Supports conductivity, TDS, salinity, resistivity, and temperature in a single instrument, reducing the need for multiple dedicated meters in water quality and environmental sampling applications.
- Automatic range segmentation: Conductivity range automatically segments from 0-19.99 µS/cm to 20.00-199.9 mS/cm, enabling precise readings across low to high conductivity samples without manual range switching.
Trade-offs
- Single-point calibration limitation: Conductivity calibration is limited to 1 point, which may reduce accuracy across wide measurement ranges compared to multi-point calibration systems, particularly in high-precision laboratory workflows.
- External power dependency: Requires a DC12V power supply, limiting portability and necessitating access to mains power or a compatible external battery source for field or remote deployment.