Description
Mini H100 Mini Dry BathProduct Type: Temperature control instrument
Research-grade temperature control platform
|
|||||||||||||||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||||||||||||||
|
TAILORED SOLUTIONS FOR RESEARCH
For bulk orders, OEM requirements, accessory matching or configuration confirmation, contact our engineering sales team.
EMAIL: inquiry@atomfair.com
|
|||||||||||||||||||||||||||||||||||||||||||||||||||
|
Supplier: Atomfair
Brand: ATOMFAIR®
|
The Mini H100 Mini Dry Bath requires operation within 10-40°C ambient temperature and below 80% relative humidity to maintain specified accuracy. Block or module compatibility must be confirmed prior to use to ensure proper sample vessel fit and heat transfer.
- Ambient Operating Conditions: Operate the dry bath within 10-40°C ambient temperature and below 80% relative humidity to prevent condensation and maintain temperature accuracy.
- Block Compatibility: Confirm that the heating block or module is compatible with the sample vessel format and fits securely on the heating platform.
- Power Supply Requirements: Connect the instrument only to a 100-240V, 50/60Hz power source using the provided DC adapter to avoid electrical damage.
- User Calibration: Use the user calibration mode to adjust temperature offset if the displayed temperature deviates from a reference thermometer.
- Error Code Diagnostics: Refer to the error code display to identify sensor or heating faults and contact technical support if needed.
This procedure describes the standard steps to set up and run a heating or incubation protocol using the Mini H100 dry bath. Ensure the instrument is placed on a stable, level surface with adequate ventilation.
Required Equipment: Mini H100 Mini Dry Bath, Compatible heating block or module, Sample vessels
- Set Temperature
Press the temperature control button and adjust to the desired set point within the range of 25 to 100 °C. - Set Timer
Select the timer mode (minutes or seconds) and enter the required incubation duration up to 999 units. - Install Block
Place the compatible heating block or module onto the heating platform ensuring full contact. - Load Samples
Insert sample vessels into the block wells and close any lid if provided. - Start Heating
Press the start button to begin the heating cycle and monitor the display for temperature and time progress.
How does the heating rate and temperature accuracy of the Mini H100 dry bath affect performance for rapid thermal cycling applications?
The Mini H100 achieves a maximum heating rate of 6.5 °C/min with a temperature control accuracy of ±0.5 °C. This combination supports repeatable incubation and enzyme reactions but may not be optimal for workflows requiring ultra-precise ramping or sub-0.5 °C uniformity across the block.
Which sample blocks or modules are confirmed compatible with the Mini H100 Mini Dry Bath?
The product description does not list specific compatible blocks or modules. It explicitly states to confirm compatible heating blocks, modules or sample vessels before ordering, and to contact the Atomfair technical sales team for module matching requirements.
What are the power supply and environmental requirements for operating the Mini H100 dry bath?
The Mini H100 accepts 100–240 V, 50/60 Hz input and requires a DC 12 V power supply with a maximum power consumption of 60 W. It must be used in a working temperature range of 10–40 °C and relative humidity below 80% RH.
This dry bath provides stable block-based heating from ambient +5°C to 100°C with ±0.5°C accuracy and a rapid heating rate of 6.5°C/min, suitable for molecular biology workflows. However, it lacks cooling ability and requires the user to source compatible heating blocks separately.
Positive
- Precise Temperature Control and Fast Heating: Temperature control accuracy of ±0.5 °C and a maximum heating rate of 6.5 °C/min ensure rapid, repeatable incubation for time-sensitive enzyme reactions and nucleic acid workflows.
- Compact and Lightweight Design: With dimensions of 110 x 162 x 140 mm and weight under 1 kg, the instrument occupies minimal benchtop space and is easily portable between lab stations.
Trade-offs
- Heating-Only Operation with No Cooling: The Mini H100 operates only above room temperature (+5 to 100 °C); it does not provide active cooling below ambient, limiting its use to warming and incubation applications.
- Requires Separate Compatible Modules: Heating blocks or sample modules are not included and must be confirmed for compatibility before ordering, adding a procurement step and potential lead time for specific vessel formats.
Every advanced material, component, equipment, and instrument in our catalog is backed by rigorous testing. We maintain strict internal quality management frameworks and align with CE conformity metrics to deliver transparent, reproducible performance data via our public open-science repository.
To request raw batch performance data, submit formal vendor registration paperwork, or execute a fast-turnaround R&D manufacturing loop, contact us at inquiry@atomfair.com.
Item is dispatched under the Atomfair Shipping & Delivery Framework (Free worldwide shipping on orders over $59 USD). Return is governed by the Atomfair Return & Refund Policy (7-day technical return window).