ATOMFAIR® EB210G-0.1MG-R-02 DUAL-RANGE ELECTROMAGNETIC ANALYTICAL BALANCERESEARCH GRADE
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
TAILORED SOLUTIONS FOR RESEARCH
Contact our engineering team for technical support or official institutional quotations.
EMAIL: inquiry@atomfair.com
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Manufacturer: Atomfair LLC
Brand: ATOMFAIR®
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
This balance requires a 30-minute pre-heating phase and sealed draft shield for 0.01 mg measurements. It features overload/underload warnings and automatic temperature compensation for accurate weighing.
- Pre-Heating Requirement: The balance must be allowed a 30-minute pre-heating phase before initializing calibration sequences to ensure electromagnetic core equilibrium.
- Draft Shield Sealing: The three-door draft shield must be entirely sealed when performing 0.01 mg readability weighing routines.
- Overload and Underload Protection: The balance emits acoustic warnings to prevent operation under overload or underload conditions.
- Leveling Requirement: The balance should be leveled using the front-mounted spirit level and adjustable leveling feet for accurate measurements.
- Temperature Compensation: The balance incorporates automatic temperature compensation to maintain accuracy across ambient temperature fluctuations.
To achieve 0.01 mg readability, seal the draft shield and allow the balance to warm up for 30 minutes. Then initialize the internal calibration sequence.
- Seal the Draft Shield
Ensure the three-door glass draft shield is entirely closed and sealed before performing semi-micro weighing routines. - Warm Up the Balance
Allow the balance to warm up for 30 minutes to let the electromagnetic core reach full equilibrium. - Initialize Calibration
Initialize the internal calibration sequence after the pre-heating phase is complete.
What is the advantage of the dual-range design on the EB210G-0.1MG-R-02 analytical balance?
The dual-range design provides two operating regimes: a 210 g maximum capacity with 0.1 mg readability for general weighing, and a finer 82 g range with 0.01 mg readability for semi-micro precision tasks. This allows the balance to handle both larger samples and high-accuracy small-mass measurements without requiring a separate instrument.
Does the EB210G-0.1MG-R-02 support automated calibration without external weights?
Yes, this model is equipped with internal calibration (FA-series variant) featuring one-key fully automated internal calibration triggered by timed intervals or ambient temperature fluctuations. This eliminates the need for manual standard weights and compensates for mechanical drift automatically.
What environmental conditions are required for achieving 0.01 mg readability on the EB210G-0.1MG-R-02?
For 0.01 mg semi-micro weighing, the three-door glass draft shield must be entirely sealed to isolate the sample chamber from air currents. Additionally, a 30-minute system pre-heating phase is required to allow the electromagnetic force compensation sensor to reach full thermal equilibrium before initiating calibration sequences.
The EB210G-0.1MG-R-02 dual-range analytical balance delivers 0.01 mg readability up to 82 g and 0.1 mg up to 210 g, with automated internal calibration via an electromagnetic force compensation sensor. However, achieving 0.01 mg precision requires a 30-minute warm-up and a fully sealed draft shield, limiting throughput.
Positive
- Dual-Range High Precision: Provides 0.01 mg readability up to 82 g and 0.1 mg up to 210 g, enabling both micro-weighing and routine analytical tasks on a single instrument.
- Internal Calibration with EMFC Sensor: Automated internal calibration eliminates manual standard weight handling, while the electromagnetic force compensation sensor ensures stable, high-resolution measurement.
Trade-offs
- Extended Pre-Heating Required: For 0.01 mg readings, a 30-minute warm-up period is necessary to stabilize the electromagnetic core, delaying immediate use.
- Limited High-Precision Range: The 0.01 mg readability is only available up to 82 g; above that, precision drops to 0.1 mg, restricting high-resolution weighing to smaller samples.