ATOMFAIR® EB210G-0.1MG-R-05 DUAL-RANGE ELECTROMAGNETIC ANALYTICAL BALANCEEXTERNAL CALIBRATION
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
TAILORED SOLUTIONS FOR RESEARCH
Contact our engineering team for technical support or official institutional quotations.
EMAIL: inquiry@atomfair.com
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Manufacturer: Atomfair LLC
Brand: ATOMFAIR®
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
This analytical balance requires a sealed draft shield and a 30-minute warm-up period when operating at 0.01 mg readability. The electromagnetic sensor must reach thermal equilibrium before calibration to ensure measurement accuracy.
- Draft Seal Requirement: Ensure the three-door draft shield housing is entirely sealed for semi-micro weighing routines at 0.01 mg readability.
- Pre-Heat Period: Allow a 30-minute system pre-heating phase to let the electromagnetic core reach full equilibrium prior to initializing calibration sequences.
What is the effective readability when using the 210 g range versus the 52 g range on the EB210G-0.1MG-R-05?
The EB210G-0.1MG-R-05 offers a dual-range matrix: 210 g maximum capacity with 0.1 mg readability and 52 g maximum capacity with 0.01 mg readability. Repeatability and linearity are ±0.6 mg for the 210 g range and ±0.06 mg for the 52 g range, enabling high-precision measurements up to 52 g at 0.01 mg resolution while maintaining 0.1 mg resolution for larger samples up to 210 g.
Can the EB210G-0.1MG-R-05 be integrated with automated data collection systems?
Yes, the balance is equipped with a standard RS232 digital output interface that supports automated computer logging loops and direct peripheral lab report printing. This allows seamless integration with laboratory information management systems (LIMS) or data acquisition software for continuous data capture.
What environmental conditions are required for achieving 0.01 mg readability on the EB210G-0.1MG-R-05?
For 0.01 mg semi-micro weighing, the three-door glass draft shield must be fully sealed to isolate the sample chamber from air currents. Additionally, a 30-minute system pre-heating phase is required to allow the electromagnetic force compensation sensor to reach full thermal equilibrium before initiating calibration sequences.
The Atomfair EB210G-0.1MG-R-05 is a dual-range analytical balance offering 210 g / 0.1 mg and 52 g / 0.01 mg capacity-readability pairs, built on a rear-mounted electromagnetic force compensation sensor with a three-door glass draft shield. External calibration is required, and achieving 0.01 mg precision demands a 30-minute warm-up and fully sealed draft shield.
Positive
- Dual-range weighing capacity: Provides 210 g / 0.1 mg and 52 g / 0.01 mg ranges, enabling both standard analytical and semi-micro precision measurements in a single instrument without switching balances.
- Rear-mounted EMFC sensor stability: High-sensitivity electromagnetic force compensation sensor rear-mounted for structural stability and resolution, supporting repeatability within ±0.6 mg (0.1 mg range) and ±0.06 mg (0.01 mg range).
Trade-offs
- External calibration requirement: Manual external calibration is required; lacks the automated internal calibration of companion FA models, demanding user verification and calibration weights for traceable linearity.
- Operational preconditioning time: For 0.01 mg readings, a 30-minute instrument warm-up is required to stabilize the electromagnetic core, and the draft shield must remain sealed against ambient air currents during measurement.
Every advanced material, component, equipment, and instrument in our catalog is backed by rigorous testing. We maintain strict internal quality management frameworks and align with CE conformity metrics to deliver transparent, reproducible performance data via our public open-science repository.
To request raw batch performance data, submit formal vendor registration paperwork, or execute a fast-turnaround R&D manufacturing loop, contact us at inquiry@atomfair.com.
Item is dispatched under the Atomfair Shipping & Delivery Framework (Free worldwide shipping on orders over $59 USD excl. heavy equipment). Return is governed by the Atomfair Return & Refund Policy (7-day technical return window).