Description
ATOMFAIR® EB82G-01MG-R-01 ANALYTICAL & SEMI-MICRO BALANCEEB82G-01MG-R-01 CORE SPECIFICATION PLATFORM
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
TAILORED INDUSTRIAL LAB SOLUTIONS
Contact our elite laboratory engineering team for institutional bulk procurement quotes or interface custom integration.
EMAIL: INQUIRY@ATOMFAIR.COM
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
MANUFACTURER OPERATIONS: ATOMFAIR LLC INSTRUMENTS DIVISION
REGISTERED BRAND: ATOMFAIR®
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
The balance requires a stable, vibration-free environment with ambient temperature between 10°C and 33°C and relative humidity between 20% and 80%. A mandatory warm-up period of 30 to 60 minutes after power-on is necessary before performing zeroing, calibration, or sample measurements.
- Temperature Constraint: Operate within 10°C to 33°C to maintain specified accuracy and repeatability.
- Humidity Constraint: Maintain relative humidity between 20% and 80% to prevent condensation and electronic drift.
- Warm-up Requirement: Allow the balance to warm up for 30–60 minutes after power engagement before any measurement or calibration.
- Calibration Trigger: Internal calibration is automatically initiated when ambient temperature changes by ≥1.5°C or when two hours have elapsed since last calibration; manual override is available.
- Vibration Isolation: Use the split-body architecture to isolate the weighing cell from hand-induced vibrations; place the display unit away from the balance base.
This procedure guides the user through initial placement, warm-up, and calibration to ensure accurate mass measurement. Following these steps minimizes drift and maximizes repeatability.
Required Equipment: Internal calibration weight (integrated), Draft shield (transparent glass)
- Position the balance
Place the balance base on a rigid, level surface away from drafts, vibrations, and direct sunlight. - Warm up the instrument
Connect the power adapter and allow the balance to warm up for at least 30 minutes (60 minutes recommended). - Perform internal calibration
Initiate the internal calibration sequence via the control panel or allow automatic calibration to occur when triggered by temperature or time. - Zero the display
Press the tare or zero button to zero the display with the empty weighing pan. - Measure the sample
Carefully place the sample on the pan and read the stable mass after the stabilization indicator appears.
How do the readability and capacity trade off between the two operating ranges of the EB82G-01MG-R-01 balance?
In the fine range, readability is 0.01 mg with a maximum capacity of 82 g, repeatability of ±0.02 mg, and linearity of ±0.02 mg. In the coarse range, readability is 0.1 mg with capacity extended to 120 g, repeatability of ±0.1 mg, and linearity of ±0.1 mg. The balance automatically switches between these dual-range matrices based on load.
What environmental conditions must be maintained for the EB82G-01MG-R-01 to achieve its specified accuracy and stability?
Operating temperature must be between 10°C and 33°C, and relative humidity between 20% and 80% RH. The balance's internal isbl calibration automatically executes if the ambient temperature fluctuates by ≥1.5°C or if more than 2 hours have elapsed, ensuring compensation for environmental drift.
Why is a 30–60 minute warm-up period mandatory before using the EB82G-01MG-R-01 for precise measurements?
The mandatory preheating threshold of 30 to 60 minutes following power-on initialization is required to allow the internal electromagnetic force sensor and electronics to thermally stabilize. This eliminates measurement drift and maintains sensor linearity, which is critical for achieving the specified repeatability of ±0.02 mg in the fine range.
The EB82G-01MG-R-01 analytical and semi-micro balance delivers dual-range readability of 0.01 mg and 0.1 mg with an internal automated calibration system (isbl) that activates upon temperature changes ≥1.5°C or after 2 hours, ensuring drift-free operation. Its split-body architecture isolates user-induced vibration, while the rear-mounted electromagnetic sensor and ≤5-second stabilization support high-throughput precision weighing in controlled laboratory environments.
Positive
- Dual-range high precision: Provides two readability levels (0.01 mg and 0.1 mg) with corresponding maximum capacities of 82 g and 120 g, allowing single-instrument adaptation to both analytical and semi-micro weighing tasks.
- Internal automated calibration: The isbl system automates recalibration when ambient temperature shifts by ≥1.5°C or after 2 hours of operation, maintaining measurement traceability without manual intervention.
Trade-offs
- Mandatory warm-up period: Requires 30 to 60 minutes of power-on stabilization before use to achieve thermal equilibrium within the sensor module; premature operation may compromise linearity and repeatability.
- Environmental sensitivity: Operates reliably only within 10°C to 33°C and 20% to 80% RH; fluctuations outside this band can degrade measurement stability and trigger frequent recalibration cycles.
Every advanced material, component, equipment, and instrument in our catalog is backed by rigorous testing. We maintain strict internal quality management frameworks and align with CE conformity metrics to deliver transparent, reproducible performance data via our public open-science repository.
To request raw batch performance data, submit formal vendor registration paperwork, or execute a fast-turnaround R&D manufacturing loop, contact us at inquiry@atomfair.com.
Item is dispatched under the Atomfair Shipping & Delivery Framework (Free worldwide shipping on orders over $59 USD). Return is governed by the Atomfair Return & Refund Policy (7-day technical return window).