ATOMFAIR® EB410G-0.001G-02 HIGH-PRECISION ELECTRONIC BALANCELABORATORY INSTRUMENT GRADE
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
TAILORED METROLOGY SOLUTIONS FOR RESEARCH
EMAIL: inquiry@atomfair.com
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Manufacturer: Atomfair LLC
Brand: ATOMFAIR®
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
This precision balance requires specific environmental and handling conditions for accurate operation. Failure to observe these constraints may compromise measurement integrity or damage the instrument.
- Environmental Vibration Isolation: Operate the balance on a stable, vibration-free bench to prevent measurement errors.
- Leveling Requirement: Level the balance using the adjustable feet and built-in level indicator before each use.
- Transport Lock Engagement: Engage the transport security lock on the glass door before relocating the balance to prevent permanent structural cell degradation.
- Draft Shield Usage: Keep the four-sided glass draft shield closed during weighing to block parasitic air currents.
- Calibration Verification: Perform calibration with a certified weight after leveling and whenever the balance is moved.
Follow these steps to safely set up and use the balance for accurate weighing. Proper initialization ensures reliable measurements and long instrument life.
Required Equipment: Calibration weight set (optional)
- Place on stable surface
Place the balance on a stable, vibration-free workbench away from drafts and heat sources. - Level the balance
Adjust the leveling feet until the bubble level is centered within the indicator. - Disengage transport lock
Verify the transport security lock on the glass door is disengaged before powering on. - Power on and warm up
Press the power button to turn on the balance and allow it to warm up for at least 10 minutes. - Perform calibration
Perform a calibration cycle using the internal calibration function or an external certified weight. - Tare the balance
Place an empty container on the pan and press the tare button to zero the reading. - Weigh the sample
Add the sample to the container and record the stable weight reading displayed on the screen.
How does the 1 mg readability of the EB410G-0.001G-02 balance interact with its ±2 mg linearity error when performing piece counting or percentage weighing applications?
The balance offers 1 mg readability, but linearity error is ±2 mg, meaning that over the full 410 g range, the deviation from true value can be up to 2 mg. For piece counting, this linearity error directly affects the calculated average piece weight, especially when counting small, lightweight items. The instrument's adjustable stabilization tracking speeds allow users to optimize response time versus accuracy, partially mitigating the impact in stable environments.
What conditions are required to achieve the stated ±1 mg repeatability on the EB410G-0.001G-02 when used for below-balance density measurements of solid-state electrolyte pellets?
The balance must be placed on a level, vibration-free surface and calibrated before use. The four-sided glass draft shield and under-pan airflow breeze break are essential to eliminate air currents that can disrupt below-balance measurements. The built-in underhook supports density setups, but the transit security lock must be disengaged and the balance properly leveled to maintain the ±1 mg repeatability specification.
What specific steps should be taken to secure the EB410G-0.001G-02 balance's internal components during laboratory relocation or storage?
The instrument includes a mechanical transit security lock on the glass door that must be engaged before any movement. Additionally, the transport security lock should be secured to prevent permanent structural cell degradation, as stated in the operational notice. The balance should be transported in its original box with protective shields to safeguard the glass and load cell.
The ATOMFAIR EB410G-0.001G-02 delivers 1 mg readability and 410 g capacity with a four-sided glass draft shield and aluminum base for environmental isolation, but demands a vibration-free setup and mandatory transport lock before relocation to avoid sensor damage.
Positive
- Environmental draft isolation: Four-sided transparent glass draft shield with aluminum alloy backplate and under-pan airflow breeze break minimizes parasitic air currents, improving measurement stability in open lab environments.
- Vibration-resistant aluminum base: Solid full-aluminum base with 4-level shock resistance and adjustable stabilization tracking speeds suppresses low-frequency ambient vibrations for consistent readings.
Trade-offs
- Vibration-free installation required: The precision apparatus must be leveled and calibrated in a vibration-free environment prior to deployment, limiting placement options without active vibration isolation.
- Mandatory transport lock during relocation: The mechanical transit security lock must be engaged before moving the balance to prevent permanent structural cell degradation, adding a procedural step for reconfiguration.