Description
ATOMFAIR® EB210G-0.1MG-R-01 DUAL-RANGE ELECTROMAGNETIC ANALYTICAL BALANCEEXTERNAL CALIBRATION
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
TAILORED SOLUTIONS FOR RESEARCH
Contact our engineering team for technical support or official institutional quotations.
EMAIL: inquiry@atomfair.com
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Manufacturer: Atomfair LLC
Brand: ATOMFAIR®
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
This analytical balance requires a 30-minute pre-heating phase to achieve electromagnetic core equilibrium before calibration. The three-door draft shield must be entirely sealed when operating in the 0.01 mg resolution mode to prevent ambient drift.
- Leveling Requirement: The balance must be leveled using the front bubble spirit level and micro-threaded leveling feet before use.
- Pre-Heating Requirement: A 30-minute system pre-heating phase is mandatory prior to initializing calibration sequences.
- Draft Shield Sealing: The three-door draft shield housing must be entirely sealed for hundred-thousandth weighing verification routines.
- Calibration Protocol: Manual external calibration must be performed to verify linearity points.
- Temperature Compensation: Integrated temperature compensation matrices maintain accuracy under varying ambient conditions.
Follow these steps to prepare the balance for accurate weighing, including leveling, pre-heating, and performing external calibration. Proper execution ensures optimal performance in dual-resolution modes.
Required Equipment: External calibration weight set
- Level the balance
Verify the balance is level using the front-facing bubble spirit level and adjust the micro-threaded leveling feet as needed. - Pre-heat the balance
Allow the balance to pre-heat for 30 minutes to stabilize the electromagnetic core before any calibration. - Seal the draft shield
Ensure all three doors of the draft shield are closed and sealed when operating in the 0.01 mg resolution mode. - Perform external calibration
Execute the manual external calibration routine using a standard weight set to verify linearity points.
How does the dual-range weighing capability of the EB210G-0.1MG-R-01 affect repeatability and linearity when switching between 210 g and 82 g capacity modes?
The balance offers a capacity-versus-precision trade-off. At the 210 g maximum capacity, repeatability and linearity are ±0.6 mg; when operating in the 82 g mode, these improve to ±0.06 mg. This allows researchers to select higher precision for smaller samples without sacrificing the ability to weigh larger masses up to 210 g.
Can the EB210G-0.1MG-R-01 be integrated into an automated data logging system for traceability in pharmaceutical quality control?
Yes, the balance includes a standard RS232 digital output interface that supports automated computer logging loops or peripheral lab report printing. Its manual external calibration protocol provides predictable, user-verified linearity points, which can be documented for audit trails in regulated environments.
What environmental precautions and warm-up procedures are required for achieving 0.01 mg readability on the EB210G-0.1MG-R-01?
For stable 0.01 mg readings, the three-door draft shield must be completely sealed to isolate the chamber from drafts and temperature fluctuations. A mandatory 30-minute pre-heating phase is required before calibration to allow the electromagnetic force compensation sensor to reach full thermal equilibrium. The balance also features structural temperature compensation and a front-mounted bubble spirit level with heavy-duty leveling feet for proper installation.
The ATOMFAIR EB210G-0.1MG-R-01 dual-range analytical balance delivers 0.1 mg/0.01 mg readability via rear-mounted EMFC sensor, with a 210 g/82 g dual capacity and RS232 output, but requires operator-performed external calibration and a 30-minute warm-up for semi-micro resolution.
Positive
- Dual-Resolution Electromagnetic Force Compensation: The rear-mounted EMFC sensor enables seamless switching between 0.1 mg and 0.01 mg readability, providing high-resolution weighing for analytical workflows while maintaining a 210 g maximum capacity.
- Comprehensive Environmental Safeguards: Built-in structural temperature compensation, overload/underload warnings, and a three-door draft shield mitigate ambient drift and physical disturbances during precision measurements.
Trade-offs
- Manual External Calibration Required: Unlike automated internal calibration variants, this FB series balance requires user-initiated manual calibration with external weights, increasing operator dependency for maintaining linearity.
- Extended Warm-Up for High-Resolution Mode: To achieve 0.01 mg readability, the balance demands a 30-minute pre-heating phase and a fully sealed draft shield, adding setup time and environmental discipline.
Every advanced material, component, equipment, and instrument in our catalog is backed by rigorous testing. We maintain strict internal quality management frameworks and align with CE conformity metrics to deliver transparent, reproducible performance data via our public open-science repository.
To request raw batch performance data, submit formal vendor registration paperwork, or execute a fast-turnaround R&D manufacturing loop, contact us at inquiry@atomfair.com.