Deep-level transient spectroscopy (DLTS) is a critical tool for analyzing defects and traps in semiconductors. The technique measures capacitance or current transients to extract defect parameters such as activation energy, capture cross-section, and defect concentration. Interpreting DLTS datasets requires robust statistical methods to ensure accuracy, particularly when dealing with overlapping signals or low signal-to-noise ratios. Key approaches include modeling defect distributions as Gaussian functions, quantifying uncertainties in activation energy calculations, and deconvolving multiple trap contributions.

Gaussian defect distributions are commonly used to represent the energy levels of traps within a semiconductor. Defects do not always exhibit a single discrete energy level but may instead have a distribution due to local strain, compositional variations, or structural disorder. A Gaussian distribution is characterized by its mean energy level (E₀) and standard deviation (σ), which describes the broadening of the defect state. The DLTS signal for such a distribution is a weighted integral over the Gaussian function, convolved with the instrument's response. Fitting the DLTS spectrum to a Gaussian model involves minimizing the residual between experimental data and the theoretical curve. The Levenberg-Marquardt algorithm is often employed for nonlinear least-squares fitting due to its stability in handling multi-parameter optimizations. The quality of the fit is assessed using the reduced chi-squared statistic, where values close to 1 indicate a good fit.

Error propagation in activation energy calculations is essential for determining the reliability of extracted defect parameters. The Arrhenius plot, which relates the emission rate (eₙ) to temperature (T), is used to derive the activation energy (Eₐ) via the equation:
eₙ = γT² exp(-Eₐ/kT),
where γ is a prefactor and k is Boltzmann’s constant. Errors in the measured emission rates and temperature uncertainties propagate into the activation energy estimate. The standard error in Eₐ is calculated using partial derivatives of the Arrhenius equation with respect to each variable. Monte Carlo simulations can further validate these uncertainties by generating synthetic datasets with added noise and comparing the resulting distributions of fitted parameters. Studies have shown that temperature uncertainties as small as ±0.1 K can lead to Eₐ errors of ±0.01 eV for deep-level defects, emphasizing the need for precise temperature control and calibration.

Multi-trap deconvolution algorithms are necessary when DLTS spectra contain overlapping peaks from multiple defects. The most common approach is peak fitting using a sum of exponential decays, each representing a distinct trap. The following steps are typically involved:
1. Initial estimation of peak positions and amplitudes via derivative or lock-in correlation techniques.
2. Simultaneous fitting of multiple exponentials to the transient data, adjusting emission rates and amplitudes.
3. Validation through residual analysis to ensure no systematic deviations remain.

For closely spaced traps, the inverse Laplace transform (ILT) method can improve resolution by converting the transient signal into a spectrum of emission rates. However, ILT is sensitive to noise, requiring regularization techniques to avoid artificial peaks. Alternative methods include the Tikhonov regularization approach, which imposes smoothness constraints on the solution. A comparison of deconvolution techniques shows that iterative fitting methods generally outperform direct inversion when the signal-to-noise ratio is below 20 dB.

A critical consideration in DLTS analysis is the impact of non-ideal conditions, such as electric field effects or trap filling inefficiencies. High electric fields can enhance emission rates via the Poole-Frenkel effect, leading to an underestimation of Eₐ if unaccounted for. Corrections involve measuring DLTS spectra at varying bias voltages and extrapolating to zero field. Similarly, incomplete trap filling due to short filling pulses introduces errors in defect concentration calculations. The filling pulse width must be optimized to ensure steady-state occupancy while avoiding excessive carrier injection.

The table below summarizes key statistical parameters in DLTS analysis:

| Parameter | Symbol | Typical Uncertainty | Impact on Results |
|--------------------|--------|---------------------|----------------------------------|
| Activation Energy | Eₐ | ±0.01–0.05 eV | Defect identification accuracy |
| Capture Cross-Section | σₙ | ±10–30% | Recombination lifetime estimates |
| Defect Concentration | Nₜ | ±5–15% | Device reliability predictions |

Advanced DLTS techniques, such as high-resolution Laplace DLTS, further refine defect analysis by resolving closely spaced energy levels. These methods rely on precise temperature scanning and sophisticated signal processing to achieve energy resolution better than 0.005 eV. However, they demand stringent noise suppression and baseline correction to avoid artifacts.

In summary, statistical methods for interpreting DLTS datasets involve Gaussian defect modeling, rigorous error propagation analysis, and multi-peak deconvolution algorithms. These techniques ensure accurate defect characterization, which is vital for semiconductor device optimization and failure analysis. Future developments may focus on enhancing computational efficiency for real-time DLTS processing, but traditional statistical approaches remain foundational for reliable data interpretation.