Atomfair Brainwave Hub: Battery Manufacturing Equipment and Instrument / Battery Testing and Characterization Instruments / X-ray Diffraction (XRD) for Material Analysis
X-ray diffraction (XRD) is a critical analytical technique for characterizing battery materials, providing insights into crystal structure, phase composition, and crystallinity. Standardized protocols, such as those established by ASTM and ISO, ensure consistency and reproducibility in XRD analysis across laboratories. This article outlines the standardized procedures for sample preparation, scan parameters, and data reporting, along with inter-laboratory validation studies.

Sample preparation is a foundational step in XRD analysis, as improper handling can introduce artifacts or preferred orientation. ASTM E915 and ISO 20203 provide guidelines for preparing powdered battery materials. The sample must be finely ground to minimize particle size effects, typically to less than 10 micrometers. For cathode materials like lithium nickel manganese cobalt oxide (NMC) or anode materials like graphite, homogeneous mixing with an inert standard such as silicon or alumina is recommended to correct for instrumental aberrations. The powder is then packed into a sample holder to ensure a flat surface, avoiding over-pressing to prevent preferred orientation. Thin-film electrodes require careful mounting to avoid substrate interference, with protocols recommending the use of low-background holders.

Scan parameters must be optimized to balance resolution, intensity, and measurement time. ASTM E390 and ISO 19950 outline standard conditions for battery materials. A typical scan range for lithium-ion battery materials is 10 to 80 degrees 2θ, using Cu Kα radiation (λ = 1.5406 Å) for most applications. A step size of 0.02 degrees and a counting time of 1-2 seconds per step are common for routine analysis. For high-resolution studies, such as detecting minor phases or lattice parameter refinement, a step size of 0.01 degrees and longer counting times may be employed. Divergence and anti-scatter slits are adjusted to optimize intensity while minimizing background noise, with guidelines specifying slit widths based on the instrument geometry.

Data reporting requirements ensure that results are comparable across studies. ASTM E2865 and ISO 19950 mandate the inclusion of instrumental parameters (e.g., radiation source, voltage, current), scan conditions (e.g., step size, counting time), and sample preparation details. For phase identification, the International Centre for Diffraction Data (ICDD) PDF-4+ database is the standard reference, with matches reported as peak positions and relative intensities. Quantitative phase analysis using Rietveld refinement requires reporting the goodness-of-fit parameters (e.g., Rwp, Rp) and the estimated standard deviations of phase percentages. Crystallite size and microstrain analysis, calculated using the Scherrer equation or Williamson-Hall method, must include the full width at half maximum (FWHM) values and the instrumental broadening correction.

Inter-laboratory validation studies are essential for establishing the reliability of XRD methods. The International Union of Crystallography (IUCr) and ASTM have conducted round-robin tests for battery materials, focusing on reproducibility of phase identification and lattice parameter measurements. In one study involving NMC cathodes, laboratories using standardized protocols achieved lattice parameter agreement within 0.01 Å for the a-axis and 0.02 Å for the c-axis. For quantitative phase analysis, deviations were typically less than 2 wt% for major phases when using Rietveld refinement. These studies highlight the importance of adherence to standardized procedures, particularly in sample preparation and instrumental calibration.

Thermal and environmental stability studies of battery materials often require non-ambient XRD. ASTM E1426 provides guidelines for high-temperature and low-temperature measurements, emphasizing the need for temperature calibration using standard materials like silver or alumina. For operando XRD studies of battery cycling, ISO 19950 recommends synchronization with electrochemical data and reporting of time-resolved diffraction patterns. The use of capillary cells or specialized sample holders minimizes air sensitivity for materials like lithium metal anodes or sulfide-based solid electrolytes.

Challenges in XRD analysis of battery materials include amorphous content detection and surface layer characterization. Standards recommend complementary techniques like pair distribution function (PDF) analysis or grazing-incidence XRD for such cases. For layered oxide cathodes, preferred orientation effects are mitigated by using spherical harmonics in Rietveld refinement or rotating sample holders during measurement.

Standardized XRD protocols are continually updated to address emerging battery materials. Recent revisions to ISO 19950 include guidelines for silicon-based anodes and high-nickel cathodes, emphasizing the need for high-resolution scans to detect subtle phase transitions. ASTM working groups are developing protocols for solid-state battery materials, focusing on interface characterization and stress measurements.

In summary, adherence to ASTM and ISO standards for XRD analysis of battery materials ensures reliable and comparable results across laboratories. Proper sample preparation, optimized scan parameters, and comprehensive data reporting are critical for accurate phase identification and quantification. Inter-laboratory studies demonstrate the effectiveness of these protocols in minimizing variability, supporting their adoption in both research and industrial settings. As battery technologies evolve, ongoing updates to these standards will address new materials and analytical challenges.
Back to X-ray Diffraction (XRD) for Material Analysis